Digital Barometers
SSD/Operating Manual 24th February 1993 PTB200-O0284-1.1
3.3 Operating commands
R Starting the measurement output
R<cr>
Command R starts the measurement output. The command is used to
start output in the STOP and RUN modes (see SMODE command) and
in interval output mode (see INTV command).
S Stopping the measurement output
S<cr>
Command S ends the RUN mode.
SEND Output a single reading
SEND<cr> STOP and RUN modes only
SEND aa<cr> POLL mode only
where
aa the address of the barometer (0...99)
SEND command is used to output one pressure reading.
See also SCOM command to define your own command for single
reading output in the STOP and RUN modes.
INTV Setting the output interval
INTV xxx yyy<cr>
where
xxx output interval (0...255)
yyy unit (s, min, h)
INTV command selects the interval output mode and sets the desired
output interval. The R command is used to start the interval outputting.