Instruction Manual

PTB 200 21 (28)
Digital Barometers
SSD/Operating Manual 24th February 1993 PTB200-O0284-1.1
RESET Resetting the barometer
RESET<cr>
resets the barometer. All software settings remain in the memory after
reset or any power failure. After changing the serial bus settings
RESET must always be given to invoke them.
ERRS Error message output
ERRS<cr>
ERRS command is used to print the error messages if any problems
occur. The command outputs an error code and error description:
>ERRS<cr>
E51 P y-value out of range
When an error occurs in the barometer, give the following seven com-
mands and print the responses on paper. On the basis of the re-
sponses to these commands Vaisala can quickly determine which ac-
tions to take. The commands are:
>ERRS<cr>
>?<cr>
>C<cr>
>F<cr>
>T<cr>
>Y<cr>
>L<cr>