Instruction Manual

PTB 200 25 (28)
Digital Barometers
SSD/Operating Manual 24th February 1993 PTB200-O0284-1.1
First precalibrate the barometer at up to eight pressure levels. Then
calculate the corrections (reference value minus PTB 200 barometer
indication) at each pressure level and enter the corrections using the
MPCI commands.
>MPCI<cr>
1. PTB reading ?499.64<cr>
correction ?0.024<cr>
2. PTB reading ?599.79<cr>
correction ?0.013<cr>
...
7. PTB reading ?1100.33<cr>
correction ?0.027<cr>
8. PTB reading ?<cr>
>MPC ON <cr>
Multip. corr. : ON
Reading Corrections
499.6400 0.0240
599.7900 0.0130
..
..
..
1100.3300 0.0270
Use ESC to abort without executing a command.
Return the MEMORY jumper to the EEPROM write disable position.