Use the following commands in the service connection to change the
flagging:
enable <Measurement Name>
disable <Measurement Name>
Examples:
>\ disable RHMeasQMH101_1
>\ disable TAMeasQMH101_1
>\ disable WMS302_1
>\ enable PWD10_1
Upon successful completion of the commands MAWS return:
Successfully disabled
or
Successfully enabled
Any other returned values indicate an error.
When the measurement input is disabled:
- All other output values than status have undefined values.
- Sensor status shows disabled, refer to the updated list of the
sensor statuses in Table 14 on page 99.
- Value status shows INVALID and NOT AVAILABLE.
where
enable = Enables the measurement inputs or sensors.
disable = Disables the measurement inputs or sensors.
Measurement Name = The measurement name from the
Configuration view of MAWS Lizard.
NOTE
For the sensors, that use more than one input channel, you need to enter
separate commands for each measurement. For example, you need to
control separately the TA and RH measurements of the QMH101
sensor.