Overall Checking
- Check signal and main cables, connectors, and connections.
- Check gaskets of the enclosures.
- Check all grounding cables, lugs, etc.
- Check mechanical assemblies, bolts, nuts, etc.
- Check for corrosion. Repair if needed.
Cable Maintenance
Inspect cables for breaks, cracks in the protective coating or cable
connectors, and bent, damaged, or misaligned pins. Also wipe off or
remove excess dirt, dust, sand, or leaves.
Updating Software to the Logger
In order to be able to utilize new features published by Vaisala, you may
need to update the logger software.
To update the software to the logger:
1. Check that you have the loader program loader.exe and the
MAWS software bin.mot on your computer in the same folder.
2. By default the loader.exe application uses the COM1 port of the
computer. When you can use the COM1 port, proceed with step a.
below, otherwise proceed as instructed in step b.
NOTE
Use the correct tools of good quality.
NOTE
Update the logger software only when requested by Vaisala.
NOTE
If any other COM port of the computer than COM1 is used for loading
the software to MAWS, you need to make a .bat file for software