User`s guide

User’s Guide ______________________________________________________________________
130 __________________________________________________________________M210662EN-A
a. Connect the maintenance terminal cable to the port COM0 of
the MAWS system and to the COM1 port on your computer.
b. Connect the maintenance terminal cable to the port COM0 of
the MAWS system and to any free COM port on your
computer. For example, if the free COM port is COM3, make
the .bat file with Notepad, see Figure 107 on page 130.
In the .bat file, the parameter -s115200 indicates the bit rate
(default is 38400) and the parameter -pCOM3 indicates the
communication port (default is COM1).
0411-044
Figure 107 Making the .bat File in Notepad
3. Close any terminal connection to MAWS so that the serial port is
free for software loading.
4. Close all other Windows programs.
5. To load the software:
a. When you are using the COM1 port of your computer, open
Command Prompt and change the directory to the folder
where you have loader.exe and bin.mot. Type loader and
press ENTER. Command Prompt window shows Waiting.
b. When you are using some other communication port, execute
the .bat file created in step 2.b. Command Prompt window
shows Waiting.
6. Reset the logger by pressing the reset button.
7. Command Prompt window shows Erasing for some time and
then starts to show Loading. If you get a load error, try again from
step 2.
8. After a few minutes, display shows Loading 100 %.
9. The program restarts with an existing setup and MAWS is now
ready to operate. If you give the long reset to MAWS, it starts the
program with a blank setup.