User`s guide
User’s Guide ______________________________________________________________________
150 __________________________________________________________________M210662EN-A
5. For adjusting the gap, place a 0.4 mm (0.016 in.) feeler gauge
between ferrite coils (parts 4 and 7) (preferably use a 30 mm wide
feeler gauge).
6. Place a screwdriver tip into the slot between the top end of rotating
transformer assembly (7) and the shaft tunnel, and pry until the air
gap between ferrite coils is 0.4 mm (0.016 in.) (use the feeler gauge
for measuring).
7. When the gap is right, carefully fasten the set screw of the rotating
transformer assembly (7).
8. Reconnect the shaft heating foil (6) and the connector for the two
white wires (3).
9. Reassemble the body cover (2) with new O-rings.
10. Tighten the three capnuts (1) and the connector nut.
11. Install the vane assembly. Align the planes in the shaft and inside
the hub. The bottom edges of the hub and the shoulder should reach
approximately the same level. Tighten the fixing screw.
The wind vane has been counter-balanced at the factory but can be
readjusted, if necessary. To do this, loosen the vane assembly and place
it on its side on the table. A correctly balanced vane will stay in
horizontal position.
CAUTION
The ferrite coils are breakable, do not try to adjust the gap by prising
them.
CAUTION
Make sure the rotating transformer assembly rotates freely without
touching the lower part at any rotary position of the shaft.
CAUTION
The heating resistance element cannot be removed without special
tools. To avoid any damages, it is recommended that replacement of
the heating element be carried out by the manufacturer.