1. For any sensor enter the following service terminal command:
LASTVAL <measurement_name> status
For example:
/ > LASTVAL TAMeasQMH101_1 status Status:1 Value:0
For details on the LASTVAL command, refer to section
LASTVAL Command on page 183.
2. For conventional sensors open the Sensor Calibration tab in
MAWS Terminal, see Figure 84 on page 96. The list of the MAWS
sensors is displayed in the tab. The Status column shows the
sensor status. To read the latest sensor statuses and measurement
values, click the Refresh button. Table 13 on page 97 describes the
other columns in the Sensor Calibration tab.
0312-038
Figure 122 MAWS Station Settings Window: Sensor Calibration
Tab
where
Status:1 = The Status field in the output is the value status.
Value:0 = The Valu e field shows the sensor status value.
CAUTION
When you upgrade a previous MAWS version to 3.06 or above, the
following sensors will not operate correctly in the Sensor Calibration
tab until they are removed from the setup and then re-created: CM6B,
CM11, QMS101, QMS102, and QMN101.