User`s guide

User’s Guide ______________________________________________________________________
184 __________________________________________________________________M210662EN-A
When the value in the report changes to ////, you can check the value. In
the following case the probe is not connected:
/ > LASTVAL TAMeasQMH101_1 raw
Status:2 Value:-238.285549
When the value is invalid, check the sensor status value, and refer to
Table 14 on page 99:
/ > LASTVAL TAMeasQMH101_1 status
Status:1 Value:0
Warnings and Errors
There may be some problems if you see either of these prompts:
/ E>
/ W>
/ E>
means that there are errors and / W> means that there are warnings.
It is normal to have one or two warnings after the serial connection to
MAWS has been reset, for example, you have turned your PC off and
so it is nothing to be worried about.
where
Status:1 = The value is valid.
Value:20.490570 = The raw value for the temperature measurement
is 20.49xxxx
where
Status:2 = The value is invalid.
Val ue: -
238.285549
= The raw value for the temperature measurement
is -238.285549, which means the probe is not
connected.
where
Status:1 = The Status field in the output is the value status.
Val ue: 0 = T he Value field shows the sensor status.