GOES Satellite Transmitter
The Satellite Data Transmitter is not serviceable in the field; if it does
not operate correctly, you must return it to Vaisala for repairs.
Ultrasonic Wind Sensor
Table 33 on page 192 lists some common problems of WS425 and their
remedies.
Combined Wind Sensor
Table 33 Troubleshooting WS425
Problem Probable Cause Remedy
When the sensor is
powered up it does not
send any data, that is,
no data is coming to the
PC.
By default, the sensor is set to a
command-response mode (polling
mode). This means that the sensor is
waiting for a data request command
from the data logger.
To change the sensor settings, you
need to open a maintenance
connection to your WS425 sensor
and configure parameter number 5,
Output Interval, to have a non-zero
value and the sensor will
automatically send data through the
serial line.
Table 34 Troubleshooting Combined Wind Sensor
Problem Probable Cause Remedy
The data is not received
from the sensor.
The sensor is mechanically
damaged.
Check the cables and connectors.
The sensor is not powered properly. Check that the supply voltage is from
3 to 15 VDC.