Chapter 4 _______________________________ Operating WMT700 in WS425 Analog Output Mode 
VAISALA _______________________________________________________________________ 33 
CHAPTER 4 
OPERATING WMT700 IN WS425 
ANALOG OUTPUT MODE 
This chapter provides information on operating WMT700 in analog 
output mode in a system configured for WS425. 
NOTE 
For general information on operating WMT700 with analog output, see 
Vaisala WINDCAP
®
 Ultrasonic Wind Sensor Series WMT700 User's 
Guide. 
Analog Output Settings 
In analog output operation WMT700 takes measurements according to 
the configured averaging time and synthesizes the analog outputs of wind 
speed and wind direction with an update interval of 0.25 seconds. The 
analog output signal type and range depend on the configured settings, 
see Vaisala WINDCAP
®
 Ultrasonic Wind Sensor Series WMT700 User's 
Guide. 
Normally analog output is enabled and the output settings are 
preconfigured at the factory according to the customer order. You can 
start using WMT700 directly after the installation with these settings. 
However, you can change the analog output settings of WMT700 at any 
time or disable the analog output functionality to save power over the 
serial interface. 
When operating WMT700 in a system configured for WS425, the analog 
output settings must be configured according to Table 10 on page 34 
using the S command. For more information on the S command, see S 
Set Parameter in Vaisala WINDCAP
®
 Ultrasonic Wind Sensor Series 
WMT700 User's Guide. 