TECHNICAL REFERENCE___________________________________________________________ 
34 __________________________________________________________________ M211097EN-C 
Table 10  Required Parameters for WS425 Analog Output 
Operation Mode 
Parameter Name  Default 
Value 
Allowed Values  Description 
aoErrValue  1000  0 …. 32000  Set 2 for voltage output and 1000 for 
frequency output for similar operation as 
with WS425. 
aout1_g 
1  0 …. 100  Gain for AOUT1. 
Set 0.01789 for compatibility with WS425 
voltage output. 
Set 11.18 for compatibility with WS425 
frequency output. 
aout2_g  1  0 …. 100  Gain for AOUT2. 
Set 0.0027 for compatibility with WS425 
potentiometer output. 
aout1mode 3  (0 = current) 
1 = voltage 
2 = frequency 
(3 = disabled) 
Analog output mode for AOUT1. 
Set 1 for voltage and 2 for frequency. 
aout2mode 7  (4 = current) 
(5 = voltage) 
6 = potentiometer 
(7 = disabled) 
Analog output mode for AOUT2. 
Set 6 for potentiometer. 
aout1_o 
aout2_o 
0  -10000 … 10000  Offset for AOUT1 and AOUT2. 
Set output offset to 0. 
For configuration instructions, see Vaisala WINDCAP
®
 Ultrasonic Wind 
Sensor Series WMT700 User's Guide. 