Chapter 5 ______________________________ Operating WMT700 with WS425 and SDI-12 profiles 
VAISALA _______________________________________________________________________ 51 
where 
<ref> = Reference: 
R = relative 
T = true 
<spd> = Wind speed 
<uni> = Wind speed unit: 
K = kilometers per hour 
M = meters per second 
N = knots 
<sta> = Status: 
A = data valid 
V = invalid data 
* = Fixed text 
<chk>  =  Checksum (8-bit XOR, excluding $ and *) 
<CR>  =  Carriage return code, ASCII 0DH 
<LF>  =  Line feed code, ASCII 0AH 
NOTE 
When the NMEA Standard profile is selected, the autoInt parameter 
must have a non-zero value since no polling command is defined for this 
profile. 
Missing Readings 
If data is missing due to a measurement problem, the NMEA messages 
show "V" in the status field. WMT700 reports 999 for wind direction and 
999.9 for wind speed. 
WS425 NMEA Extended Profile (v. 0183) 
When the WS425 A/B NMEA Extended profile is selected, you can set 
WMT700 to send messages based on the configured automatic message 
interval or poll data using the $WIP command. 
Configurable Parameters 
Table 15 on page 52 lists the configurable parameters and their allowed 
and default values for the WS425 A/B NMEA Extended profile. 