TECHNICAL REFERENCE___________________________________________________________ 
58 __________________________________________________________________ M211097EN-C 
Interpretation of the example message: 
- 
┐
 02H (<STX>, start of transmission) 
- Sensor ID: W 
-  Averaging in seconds: 5 
-  Sensor status: P = pass 
-  Wind direction: 120 
-  Wind speed: 0013 
-  . (dot character) 
-  Wind speed unit: T = meters per second 
-  Checksum (most significant digit): D 
-  Checksum (least significant digit): E 
- 
└
 03H (<ETX>, end of transmission) 
Missing Readings 
If data is missing due to a measurement problem, the WS425 A/B ASCII 
message reports 999.9 for wind speed. 
WS425 A/B WAT11 Profile 
Configurable Parameters 
Table 20 below lists the configurable parameters and their allowed and 
default values for the WS425 A/B WAT11 profile. 
Table 20  Configurable Parameters for WS425 A/B WAT11 
Profile 
Parameter Default 
Value 
Profile-specific Allowed 
Values 
Description 
com1_protocol 
com2_protocol 
(depending on 
the port used 
for operating 
WMT700) 
0  6 = WS425 A/B WAT11 
Profile for serial ports COM1 and COM2. 
To start continuous measurement, use the START command. You can 
stop the measurement with the STOP command. For instructions, refer to 
Vaisala WINDCAP
®
 Ultrasonic Wind Sensor Series WMT700 User's 
Guide. 