Chapter 5 ______________________________ Operating WMT700 with WS425 and SDI-12 profiles 
VAISALA _______________________________________________________________________ 65 
aAb! — Change Address 
This command changes the WMT700 address. After WMT700 has 
received this command and sent a response, the sensor does not respond 
to any other command for one second. This gives WMT700 time to write 
the new address to the non-volatile memory. 
NOTE 
You can also change the WMT700 address by entering configuration 
mode and changing the address parameter. 
aAb! 
where 
a = The current single-digit WMT700 address that 
corresponds to the first character of the address value. 
A  =  The change address command. 
b  =  The new address. 
!  =  Terminates the command. 
The response is the following: 
b<CR><LF> 
where 
b  =  The new single-digit WMT700 address (or the original 
address if WMT700 is unable to change the address). 
<CR><LF>  =  Terminates the response. 
aC! — Start Concurrent Measurement 
This command starts concurrent measurement. WMT700 does not return 
measurement data after this command. It sends a response that indicates 
when the measurement result can be retrieved using the aD0! command 
and the number of parameters. 
aC! 
where 
a  =  The single-digit WMT700 address that corresponds to 
the first character of the address value. 
C  =  The start concurrent measurement command. 
!  =  Terminates the command. 