Chapter 5 ______________________________ Operating WMT700 with WS425 and SDI-12 profiles 
VAISALA _______________________________________________________________________ 69 
aV! — Start Verification 
This command starts verification. However, WMT700 does not return 
verification data directly after this command. It sends a response that 
specifies when the verification results can be retrieved using the aD0! 
command and the number of parameters. The WMT700 verification data 
includes a watchdog count, sensor measurement unit settings, and the 
current SDI mode setting. 
The command is the following: 
aV! 
where 
a  =   The single-digit WMT700 address that corresponds to the first 
character of the address value. 
V  =  The start verification command. 
!  =  Terminates the command. 
The response is the following: 
atttn<CR><LF> 
where 
a  =  The single-digit WMT700 address that corresponds to 
the first character of the address value. 
ttt  =  The period of time, in seconds, after which WMT700 has 
the verification data ready. 
n  =  The number of verification data fields that will be 
returned. 
<CR><LF> =  Terminates the response. 
Example of the command and response: 
1V! 
10014<CR><LF> 
Interpretation of the example message: 
-  Time when the measurement is ready: 1 second 
-  Number of returned measurement values: 4 
To fetch the measurement results, use the aD0! command. 