SDI-12 Data Messages 
When SDI-12 profile is selected, the data message sent by WMT700 
depends on the command that you use before fetching the data with aD0! 
command. 
WS425 A/B SDI-12 Message for C and M Command 
When you have given the commands aC! or aM! followed by aD0!, the 
response is the following: 
a<WS><WD><x><y><s><CR><LF> 
where 
a  =  The single-digit WMT700 address that corresponds 
to the first character of the address value. 
<WS>  =  Polar wind speed in selected units 
Format: +(ss)s.s 
<WD>  =  Polar wind direction in degrees 
Format: +(dd)d.d 
<x>  =  x component of wind speed in selected units 
Format: ±(ss)s.s 
<y>  =  y component of wind speed in selected units 
Format: ±(ss)s.s 
<s>  =  Static speed of sound in miles per hour 
Format: ±(ss)s.s 
<CR><LF>  =  Terminates the response. 
The above are measurement data fields. 
Missing Readings 
If data is missing due to a measurement problem (for example, blocked 
paths between transducers), the measurement data is replaced with 999.9. 
Example of the command and response when data is missing: 
7D0! 
7+999.9+999.9+999.9+999.9+999.9 <CR><LF> 