3.4.3. Performing a Quick Test
Aquicktestcanbeperformedwithouthavingtoenterthetestcurrent,theburnintime,and
theramptime.Foraquicktest,acurrentof100Aisused,andtheburnintimeandramptime
arebothsetto5seconds.TheQuickTestmodeshould
notbeusedifthecircuitbreakeris
connectedtodualgrounds(iftheunithastheDualGroundoption).
NOTE
TheQuickTestmodeshouldnotbeusedifthecircuitbreakerisconnectedtodual
groundsbecausetheAutoOhmwillnotmonitortheexternalcurrentprobecurrent.
Usethestepsbelowtoperformaquicktest:
a. Startfromthe“STARTUP”menu:
Pressthe[3]key(QUICKTEST)
b. Thefollowingscreenwillbedisplayed
Pressthe
[START]key.
c. Theunitwillstartrampingupthecurrentto100A.Thefollowingscreenwillbe
displayed:
Oncethecurrenthasreached100A,theinitialtestresultswillbedisplayedandthe
burninsequencewillbeinitiated.Thefollowingscreenwillbedisplayed:
ramping current
12.30A 27
test in progress...
test parameters
current: 100 amps
burn-in: 5 sec
ramp: 5 sec
"START" TO RUN TEST
1.RUN TEST
2.SETUP
3.QUICK TEST
time: 14:21:34
date: 05/24/11