4. For each tap connection you would like to test, click on the check box under each test
type that you would like to perform. For example, to perform a ratio test on the X1-X5
tap connection, check the check box under the “Ratio Test” column in the X1-X5 row.
a. Once a test has been selected for a tap connection, the “Nameplate Ratio” input
fields will be activated. Enter the nameplate ratio values for the tap connection.
b. Select the excitation voltage from the “Excitation Voltage” drop-down list
corresponding to the tap connection.
c. Enter any relevant test notes in the corresponding “Test Note” field.
5. Click on the “Save” button. The following window will be displayed:
Browse to the location where you would like to save the test plan. Enter a file name in the “File
name:” input field and click on the “Save” button. The test plan will be saved in the selected
location.