User guide

TRI-PHASE OPERATING™ INSTRUCTIONS
12
3.0 Functional Description (continued)
The TTRA software application is compatible with Microsoft Windows XP, and Microsoft
Windows Vista (see section
11.3 for more details).
Capable of raising or lowering the Load Tap Changer (LTC) tap position from the TRI-
PHASE™ front panel with the built-in LTC controller (see section
24.0 for more details).
4.0 Principles of Operation
The TRI-PHASE™ measures transformer turns-ratios (using the ANSI/IEEE C57.12.90 method) by
applying a test voltage across the primary (H) winding and sensing the induced voltage on the
secondary (X) side. For safety, testing is always done in a step-down transfer, regardless of the
transformer’s actual use. Since there is no load on the windings during testing, the measured voltage
ratio is virtually the same as the winding turns-ratio.
The TRI-PHASE™ checks for test cable hook-up errors before each test. It applies a low-level test
voltage (300 mV) across the winding being tested and senses the induced secondary voltage. If the
induced voltage is greater than the applied excitation voltage, a hookup error is assumed. If a
connection error is detected, the TRI-PHASE™ aborts the test and displays “Hook-Up Error” on the
LCD. If no hookup error is detected, the TRI-PHASE™ applies a full test voltage to the transformer
winding being tested and the turns-ratio (or voltage ratio) is displayed on the LCD.
The winding polarity and phase-angle are determined by comparing the induced voltage waveform
to the test voltage waveform (which is used as the reference). In-phase waveforms (+) measure a
phase angle centered about 0 degrees. Out-of-phase waveforms () measure a phase angle centered
about 180 degrees.
The TRI-PHASE™ measures turns-ratios in the range from 0.8 to 15,000. Excitation current
(flowing in the H leads) is measured for reference and ranges from 0 to 2,000mA. Winding polarity
is displayed as a “+” or “”sign in front of the measured ratio. The phase angle is measured in
degrees with a resolution of ±0.2 degrees.
NOTE:
The phase angle measurement is not displayed on the LCD.
The phase angle measurement will be printed when the detail print format is selected
and displayed with a PC running the TTRA software application.