Manual

REV 3 WRM-10P AND WRM-40 USER’S MANUAL
21
Press the [1] key (COPPER, Tk=234.5) to select copper as the winding material.
Continue to step d.
Press the [2] key (ALUMINUM, Tk=225) to select aluminum as the winding
material. Continue to step d.
Press the [3] key (MANUALLY ENTER Tk) to manually enter the Tk value. The
following screen will be displayed:
You can increase the Tk value by 0.5°C increments by pressing the [PAPER ∧
Contrast]
key.
You can decrease the Tk value by 0.5°C increments by pressing the [PAPER ∨
Contrast]
key.
Press the [ENTER] key to confirm the temperature setting. Continue to step e.
2. NO
Press the [2] key (NO) if you do NOT wish to convert readings to standard
temperature. Continue to step e.
e. The following screen will be displayed:
Use the
[PAPER ∧ Contrast] and [PAPER ∨ Contrast] keys to adjust the D.U.T.
(Device Under Test) temperature.
Press the [ENTER] key to confirm the D.U.T. temperature setting.
f. The following screen will be displayed:
Use the [PAPER ∧ Contrast] and [PAPER ∨ Contrast] keys to adjust the
reference temperature.
Press the
[ENTER] key to confirm the reference temperature setting.
REFERENCE TEMP:
85.0°C 185.0°F
↑↓ to adjust temp
“ENTER” to accept
D.U.T. TEMP
25.0°C 77.0°F
↑↓ to adjust temp
“ENTER” to accept
Tk:
230.0°C
↑↓ to adjust Tk
“ENTER” to accept