5.0 Troubleshooting Guide
PROBLEM DESCRIPTION Probable Cause Suggested Solution
No current displayed on
the LCD screen when
performing a test.
Opened current cable connection.
Opened transformer windings.
Check current cable connection.
Perform a quick test using a shunt
or a piece of aluminum bar.
Resistance reading is zero
or invalid while current
value is displayed on the
LCD screen.
Opened sense leads connection.
Sense leads problem.
Check sense leads connection to
D.U.T.
Check sense leads using an
ohmmeter.
Perform a diagnostic test. See
section 3.8.