Specifications

LA Series Linear Servo Amplifier Technical Reference Manual
27
11 Balancing
Motor phase balancing is performed to minimize torque ripple in the motor caused by impedance
variations in the motor windings and to compensate for component tolerance variations in the amplifier
drive circuits. The intent of this adjustment is to get the three motor voltages approximately equal taking
into account polarity. Once the drive is balanced, the setting should not need to be changed as long as the
drive operates with the same motor and at the same baseplate temperature as when the balance was
performed
NOTE: Be sure the amplifier is at operating temperature when performing any balancing procedures. To
achieve operating temperature, enable the amplifier with the motor load connected and monitor the
baseplate temperature, either at TP2 (10°C/Volt) or using the “L” serial command.
There are two methods that can be used to balance the output phases, Autobalance and Manual
Balance. For most cases, the Autobalance feature will balance the phases to an acceptable level. For
other cases where the motor resistance may be very high, or exact phase balance must be obtained, the
manual method should be used after the Autobalance to further “tweak” the offsets.
11.1 Autobalance
The drive has the ability to autobalance the phase offsets using an internal algorithm. The algorithm is
initiated by holding the push button for longer than 1 second. The “B” command can also be used from the
serial interface. Once started, the algorithm will display a “-“ middle bar on the LED display. Depending on
the amount of offset, the phase voltage and the load resistance, the amount of time spent in the
Autobalance routine varies from about 5 seconds to about 30 seconds.
If the drive is unable to balance the phases, the routine exits and reports and Autobalance Fault
(4) on the LED display.
Note: Be sure the motion controller is set to output 0vdc to the command input(s) and the motion controller
is set for open loop mode (no feedback loop). This is very important, as the motor will not be able to move
during this procedure.
The procedure for using the Autobalance feature:
1) Connect the motor in the normal manner.
2) Apply power to the system (Bias and Bus).
3) Set the motion controller to open loop mode (no feedback loop)
4) Set the motor so no motion is possible (lock down the stage).
5) Set the command input(s) to 0 volts and Enable the amplifier.
6) Press the Reset button S1 and hold for greater than 1 second. Confirm the middle bar on the LED
display is showing. Release the button.
7) Once the procedure completes, the display will either show Enabled (0) or Autobalance Fault (4). If
the drive is Enabled, the process has completed successfully. If a fault is shown, the phases will
have to be balanced manually.
Note: The Autobalance function is typically able to balance a load with a resistance that falls within the
range of 1 ohm to 10 ohms. The Autobalance function may not consistently balance loads outside this
range. This does not indicate a problem with the amplifier. It means the load may need to be manually
balanced. This is especially true if the manual balance pots have been moved from their factory center
position. Once the load has been manually balanced and the pots re-centered, the Autobalance function
will most likely work.