User Manual

Table Of Contents
4343W X-ray Detector
16 www.vareximaging.com
REFERENCE MANUAL
8.7.3 Battery Charge Level
The Battery charge level may be found on the Battery. Press the indicator button on the Battery
and charge level will illuminate. Each LED illuminated represents 25% charge.
Figure 11 Battery Charge Level
8.7.4 Battery Hot-Swap
The 4343W X-ray Detector is equipped with a Hot-Swappable Battery. When a discharged
Battery is removed from the X-ray Detector, the User will have approximately 3 minutes of
operation for the Battery Hot-Swap to occur.
8.8 Optional Varex Provided Inductive Battery Charger
The Battery may be charged using the Varex provided Inductive Battery Charger while the X-ray
Detector is installed in the Bucky Tray.
8.8.1 Installing the Varex Provided Inductive Battery Charger
1 Install the Varex Inductive Battery Charger and Detector Support to the Bucky Tray.
2 Self-tapping screws with washers, machine screws with washers into PEM nuts (installed on
the Bucky Tray), or VHB Tape may be used to attach the Varex Inductive Battery Charger
and Detector Support into the Bucky Tray. Use Figure 12, Figure 13, and Figure 14 for
mounting locations.
3 Route the USB-C Cable from the Varex Inductive Battery Charger to the Power Supply in a
way that the opening and closing of the Bucky Tray does not cause damage to the cable. The
USB-C Cable can be replaced if damaged.
WARNING
If a fully charged battery is not re-inserted within the time window, the
X-ray Detector will power-OFF.
Note
Acquisition and calibration features are not available while Hot-Swapping
a Battery.
Note
During image acquisition or calibration, inductive battery charging will
pause.