User Manual
Table Of Contents
- 4343W X-ray Detector
- Table of Contents
- List of Tables
- List of Figures
- 1.0 Scope
- 2.0 Contraindication
- 3.0 Intended Use
- 4.0 Audience
- 5.0 Abbreviations
- 6.0 Definition of Symbols
- 7.0 Standards and Regulations
- 8.0 Description of the 4343W
- 8.1 Shipment Contents
- 8.2 Optional Parts
- 8.3 X-ray Detector Surfaces and Features
- 8.4 X-ray Detector Dimensions
- 8.5 X-ray Detector Specifications
- 8.6 Environmental Considerations
- 8.7 X-ray Detector Battery
- 8.8 Optional Varex Provided Inductive Battery Charger
- 8.9 Designing a Custom Inductive Battery Charger
- 8.10 Using a Third-Party Inductive Battery Charger
- 8.11 Power-on and Power-off Sequence
- 8.12 Reboot Sequence
- 8.13 LED Status Indicator Behavior
- 8.14 Position Sensing
- 8.15 Detector Sharing
- 9.0 The Detector Configuration File
- 10.0 System Software
- 11.0 Maintenance
- 12.0 Troubleshooting
- 13.0 Disposing of the X-ray Detector and Battery
- 14.0 Safety - Electromagnetic Interference
- 15.0 Regulatory
- Index
4343W X-ray Detector
22 www.vareximaging.com
REFERENCE MANUAL
8.13 LED Status Indicator Behavior
Figure 17 LED Status Indicator
8.14 Position Sensing
The 4343W is equipped with two position sensors that communicate with software the X-ray
Detector’s exact position within a bucky setup. Refer to document 03236 FP-2032 - Varex Smart
Panel Software Interface for integration of Position Sensing.
Note
The blinking behavior occurs based on a 4Hz clock. Each digit for the
blinking pattern represents 1/4s. 0 =LED OFF, 1 =LED ON, X = Previous
State.
Table 8 LED Status Details
LED Behavior Status
Orange Solid (1111) Booting
Green Blinking (100000) Not connected to Wireless Router, blinks every
1.5 seconds
Green Blinking (1010) Connected to Wireless Router, blinks twice (2)
per second
Green Solid (1111) Link Opened, detector controlled remotely, LED always
on
Green Blinking (1100) Connected to Service Cable, blinks once (1) per second
Yellow Solid (1111) Detector Error
Purple Blinking (1xxxxxxx) Battery Hot-Swap Active
(battery exhausted or removed)
Blue Blinking (1xxxxxxx) Inductive Charger is Active (detector is mated with the
inductive charger and charging is enabled)
No LED Off