User Manual

Table Of Contents
www.vareximaging.com 23
4343W X-ray Detector
REFERENCE MANUAL
8.15 Detector Sharing
The 4343W X-ray Detector is equipped with Detector Sharing capability. Detector Sharing allows
the IP address, password, and SSID to be set via software command. The software command for
Detector Sharing can be sent via wireless or the Service Cable. Refer to document 03236
FP-2032 - Varex Smart Panel Software Interface for more information.
9.0 The Detector Configuration File
A configuration file will ship with the X-ray Detector. The file is loaded by Varex onto the X-ray
Detector and will contain 1 or more modes. The integration time or window exposure time is set
dynamically. Modes have a default integration time of 1000ms. Refer to document 03236
FP-2032 - Varex Smart Panel Software Interface for more information on how to dynamically
change the integration time.
The purpose of each mode is to configure the X-ray Detector to achieve optimal performance
during specific imaging procedures. Each mode is a combination of settings, such as; cycle time,
rad or vTrigger, detector type and integration time. Modes are defined in the system
configuration files. The configuration file is stored in the X-ray Detector’s non-volatile memory.
The 4343W supports Radiographic and vTrigger modes of operation, see Table 9 for a typical
mode example. The sensitivity of the X-ray Detector is optimized to match the X-ray dose used in
each mode.
Table 9 Operational Mode Example
Characteristic Specification
Modes Radiography or vTrigger
Pixel Binning 1 x 1
Integration Time 1000ms default (350ms - 4000ms possible)
X-ray Window Time 0.35 to 4 seconds Dynamic Integration
Cycle Time (@550ms) <
5 seconds
Image Area Full Field
Frame Size 3072 x 3072