4343W X-ray Detector
Caution
It is important that the X-ray Detector is not directly connected to the
installed network. Connection of the X-ray Detector with the installed
computer network may disturb the IT environment.
Caution
Do not operate the X-ray Detector in a location with the following
conditions:
Close to fluid or places where fluid is used
Close to heat sources, such as a heater
High temperature environment
High humidity environment
High condensation environment
Extreme cold environment
Dusty environment
Salty or sulfurous environment
Near a vibrating environment
Environment where there is insufficient air circulation to enable the X-ray
Detector and power supply to dissipate heat
Ignoring this caution may result in personal injury or damage to
the product.
Caution
The X-ray Detector is an applied part (patient contact device) and the
surface shall not exceed 42°C internal temperature sensor data is
provided in the diagnostic data attached to each image. These
temperature measurements are well correlated with the X-ray Detector
external surface temperature. It is advisable to monitor this diagnostic
data as an additional safety precaution, see document 03236 FP2032 -
VSP Software Interface. Also, see Table 7.
Note
The X-ray Detector is intended to be installed, maintained, and used in a
professional healthcare facility.
Caution
Monitor the temperature of the X-ray Detector and allow the X-ray
Detector to cool down when the temperature of the X-ray Detector gets
hot. The device is not designed to supply heat to a patient. Ignoring
this caution may result in personal injury or damage to the product.
Caution
An Image Quality Test or Customer Acceptance test should be
performed before the X-ray Detector is used for the first time with
patients.
No
Caution
No part of the X-ray Detector is intended to be attached to the patient
and/or contact the patient.