User Manual
Table Of Contents
- 4343W X-ray Detector
- Table of Contents
- List of Tables
- List of Figures
- 1.0 Scope
- 2.0 Contraindication
- 3.0 Intended Use
- 4.0 Audience
- 5.0 Abbreviations
- 6.0 Definition of Symbols
- 7.0 Standards and Regulations
- 8.0 Description of the 4343W
- 8.1 Shipment Contents
- 8.2 Optional Parts
- 8.3 X-ray Detector Surfaces and Features
- 8.4 X-ray Detector Dimensions
- 8.5 X-ray Detector Specifications
- 8.6 Environmental Considerations
- 8.7 X-ray Detector Battery
- 8.8 Optional Varex Provided Inductive Battery Charger
- 8.9 Designing a Custom Inductive Battery Charger
- 8.10 Using a Third-Party Inductive Battery Charger
- 8.11 Power-on and Power-off Sequence
- 8.12 Reboot Sequence
- 8.13 LED Status Indicator Behavior
- 8.14 Position Sensing
- 8.15 Detector Sharing
- 9.0 The Detector Configuration File
- 10.0 System Software
- 11.0 Maintenance
- 12.0 Troubleshooting
- 13.0 Disposing of the X-ray Detector and Battery
- 14.0 Safety - Electromagnetic Interference
- 15.0 Regulatory
- Index
REFERENCE MANUAL
www.vareximaging.com v
4343W X-ray Detector
Caution
It is important that the X-ray Detector is not directly connected to the
installed network. Connection of the X-ray Detector with the installed
computer network may disturb the IT environment.
Caution
Do not operate the X-ray Detector in a location with the following
conditions:
•
Close to fluid or places where fluid is used
• Close to heat sources, such as a heater
• High temperature environment
• High humidity environment
• High condensation environment
• Extreme cold environment
• Dusty environment
• Salty or sulfurous environment
• Near a vibrating environment
• Environment where there is insufficient air circulation to enable the X-ray
Detector and power supply to dissipate heat
Ignoring this caution may result in personal injury or damage to
the product.
Caution
The X-ray Detector is an applied part (patient contact device) and the
surface shall not exceed 42°C internal temperature sensor data is
provided in the diagnostic data attached to each image. These
temperature measurements are well correlated with the X-ray Detector
external surface temperature. It is advisable to monitor this diagnostic
data as an additional safety precaution, see document 03236 FP2032 -
VSP Software Interface. Also, see Table 7.
Note
The X-ray Detector is intended to be installed, maintained, and used in a
professional healthcare facility.
Caution
Monitor the temperature of the X-ray Detector and allow the X-ray
Detector to cool down when the temperature of the X-ray Detector gets
hot. The device is not designed to supply heat to a patient. Ignoring
this caution may result in personal injury or damage to the product.
Caution
An Image Quality Test or Customer Acceptance test should be
performed before the X-ray Detector is used for the first time with
patients.
No
Caution
No part of the X-ray Detector is intended to be attached to the patient
and/or contact the patient.