5 An Offset Calibration Progress window appears.
Figure 23 Offset Calibration Progress
6
Once all frames are acquired, the X-ray Detector is updated with the averaged offset
calibration frame for the current mode. The updated message will show on the progress
window and calibration process is complete.
7 Click Close.
Figure 24 Finish Offset Calibration