4343W X-ray Detector
REFERENCE MANUAL
10.5.4 Gain Calibration Steps
1 Choose the desired imaging mode from the Mode Drop Down Menu.
Figure 25 Mode Drop Down Menu
2
Under the Acquisition menu, click Gain Calibration. This invokes hardware handshaking
for the dark field calibration.
Figure 26 Initiate Gain Calibration
WARNING
In order to acquire an image with X-ray exposure in vTrigger mode,
Hardware Handshaking must be enabled, see Figure 34, or an image
without exposure will be returned.