User Manual

Table Of Contents
REFERENCE MANUAL
www.vareximaging.com vii
4343W X-ray Detector
Handling
WARNING
Never disassemble, modify, or alter the X-ray Detector, its
components, battery pack, battery charger, or accessories. Ignoring
this warning may cause electrical shock and/or unknown hazards,
which may result in severe personal injury, death, or substantial
product damage.
WARNING
Do not use the Battery Latch as a handle. Ignoring this warning may
cause damage to the Battery Latch or increase the likelihood that the
X-ray Detector may be dropped causing substantial product damage.
Caution
Place the X-ray Detector horizontally on a flat, stable surface. If the
X-ray Detector is placed vertically or in any tilted position, the X-ray
Detector must be securely placed in the Bucky tray or securely
fastened to the X-ray Detector enclosure or support structure. Ignoring
this caution may result in personal injury or damage to the product.
Caution
Do not exceed the maximum uniform load weight of 300 kg
distributed across the surface of the X-ray Detector.
Caution
Do not exceed the maximum load weight of 200 kg distributed on an
area of 40 mm in a diameter of the X-ray Detector surface.
Caution
Do not drop the X-ray Detector. If the X-ray Detector is dropped,
remove the X-ray Detector from service, and immediately ask your
establishment’s safety representative to verify or re-validate the
proper function of the X-ray Detector prior to resuming use of the
X-ray Detector. Further use under abnormal conditions may result in
severe personal injury, death, or substantial product damage.
Caution
In the event that the X-ray Detector is dropped, a new Gain
Calibration must be performed.
Caution
Do not expose the backside of the X-ray Detector with X-rays. Always
use the top side of the X-ray Detector for examinations.