User Manual
Table Of Contents
- 4343W X-ray Detector
- Table of Contents
- List of Tables
- List of Figures
- 1.0 Scope
- 2.0 Contraindication
- 3.0 Intended Use
- 4.0 Audience
- 5.0 Abbreviations
- 6.0 Definition of Symbols
- 7.0 Standards and Regulations
- 8.0 Description of the 4343W
- 8.1 Shipment Contents
- 8.2 Optional Parts
- 8.3 X-ray Detector Surfaces and Features
- 8.4 X-ray Detector Dimensions
- 8.5 X-ray Detector Specifications
- 8.6 Environmental Considerations
- 8.7 X-ray Detector Battery
- 8.8 Optional Varex Provided Inductive Battery Charger
- 8.9 Designing a Custom Inductive Battery Charger
- 8.10 Using a Third-Party Inductive Battery Charger
- 8.11 Power-on and Power-off Sequence
- 8.12 Reboot Sequence
- 8.13 LED Status Indicator Behavior
- 8.14 Position Sensing
- 8.15 Detector Sharing
- 9.0 The Detector Configuration File
- 10.0 System Software
- 11.0 Maintenance
- 12.0 Troubleshooting
- 13.0 Disposing of the X-ray Detector and Battery
- 14.0 Safety - Electromagnetic Interference
- 15.0 Regulatory
- Index
www.vareximaging.com 51
4343W X-ray Detector
REFERENCE MANUAL
11.1.2 Cleaning the X-ray Detector
To clean the X-ray Detector:
1 Remove the battery from the X-ray Detector.
2 Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
3 Dispense wipe from container.
4 Visually inspect the device. If contamination is present, use a wipe to remove the
contamination from the device surface. Use multiple wipes, as applicable, to remove the
visible contamination.
5 Take care to remove contamination from seams, joints, and other difficult-to-reach areas.
6 Dispose of each wipe used for cleaning.
7 If visible residue from the cleaning product is evident after air-drying, remove the residue
with a general Isopropyl Alcohol wipe, and let air-dry.
11.1.3 Disinfecting the X-ray Detector
To disinfect the X-ray Detector:
1 Remove the battery from the X-ray Detector.
2 Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
3 Wipe the entire surface of the target areas of the system until they are visibly wet with the
chemical solution. Target areas include any surface that was cleaned and/or any other
surfaces potentially contaminated during system use.
4 Take care to wet seams, joints, and other difficult-to-reach areas.
5 Ensure that the surfaces remain visibly wet for the maximum necessary disinfection time
specified on the product label.
a If a disinfection time is not specified on a chemical label for the concentration used,
ensure that the surfaces remain visibly wet for a minimum of 10 minutes.
b Use additional fresh wipes as needed to ensure continuous wet contact time during the
specified disinfection period.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.