User Manual

VEGAPULS 42 and 44 4 20 mA 9
Product description
Temperature influence: Temperature error absolutely
zero (e.g. at 500°C 0.018 %)
Pressure influence: Error with pressure increase very
low (e.g. at 50 bar 1.44 %)
100 500 1000 1300 ˚C
0
0
0,01
0,02
0,03
%
0,018 %
0,023 %
10
0
0
5
%
1,44 %
2,8 %
10
50
20 30 40 60
100
70 80 90 110 120 130 140
bar
0,29 %
3,89 %
Continuous and reliable
Unaffected by temperature, pressure and
individual gas atmospheres, VEGAPULS
radar sensors are used for quick and reliable
continuous level measurement of various
products.
VEGAPULS 40 sensors enable level meas-
urement with radar in facilities where previ-
ously, due to high cost, it was completely out
of the question.
1.2 Application features
Applications
level measurement of liquids
measurement also in vacuum
all slightly conductive materials and all
substances with a dielectric constant > 2.0
can be measured
measuring range 0 10 m (type 42).
measuring range 0 20 m (type 44).
Two-wire technology
supply and output signal on one two-wire
cable (Loop powered)
4 20 mA output signal or HART
®
output
signal.
Rugged and abrasion proof
non-contact
high-resistance materials
Exact and reliable
accuracy 0.05 %.
resolution 1 mm.
unaffected by noise, vapours, dusts, gas
compositions and inert gas stratification
unaffected by varying density and tem-
perature of the medium
measurement in pressures up to 40 bar
and product temperatures up to 200°C
Communicative
integrated display of measured value
optional display module separate from
sensor
adjustment with detachable adjustment
module, pluggable in the sensor or in the
external display
adjustment from the PLC level with the PC
adjustment with HART
®
handheld
Approvals
CENELEC, ATEX, PTB, FM, CSA, ABS,
LRS, GL, LR, FCC.