User Manual

8 Maintenance and fault rectication
8.1 Maintenance
When used as directed in normal operation, VEGAPULS
68 is completely maintenance-free.
8.2 Fault rectication
VEGAPULS 68 oers maximum reliability. Nevertheless,
faults may occur during operation. These can have the
following causes, e.g.:
l Sensor
l Process
l Power supply
l Signal processing.
The rst measures to be taken are to check the output
signal and evaluate fault messages via the adjustment
module PLICSCOM. The procedure is described below.
Further comprehensive diagnostics can be carried out
on a laptop with the software PACTware and the
suitable DTM. In many cases, the reasons can be
determined in this way and faults can be rectied.
Should these measures not be successful, please call in
urgent cases the VEGA service hotline under the phone
number +49 1805 858550.
The hotline is available to you 7 days a week round-the-
clock. Since we oer this service world-wide, the
support is only available in the English language. The
service is free of charge, only the standard telephone
costs will be charged.
Connect a hand-held multimeter with a suitable measu-
ring range acc. to the wiring plan.
? 4 ... 20 mA signal is not stable
l level uctuations
à set integration time via PLICSCOM or
PACTware
Causes of failure
Fault rectication
24 hour service hotline
Checking the 4 ... 20 mA
signal
VEGAPULS 68 - 4 ... 20 mA/HART 45
Maintenance and fault rectication
29261-EN-050202