11 Close the opening levers of the terminals, you will
hear the terminal spring closing
12 Check the hold of the wires in the terminals by lightly
pulling on them
13 Check the tightness of the cable entry. The seal ring
must completely encircle the cable.
14 Screw the housing cover back on
The electronics exchange is nished.
As a rule, the exchange of the oscillator must be
documented internally when used in Ex applications.
If the serial number of the oscillator is stated when
ordering, VEGAPULS 68 is ready for operation directly
after mounting the oscillator.
If you are using an universal electronics from stock, you
have to load the sensor data after mounting the
oscillator.Use the item "serial number search" under
www.vega.com. After having entered the serial number,
the order data of the sensor will be displayed. Below the
order data you will nd "Sensor data for Service-DTM"
as an XML le. Load this le to your PC and then into the
sensor via PACTware/Service DTM.
After VEGAPULS 68 is ready for operation, settings
carried out on site must be repeated.
Tip:
Use the copy function of the adjustment module
PLICSCOM or the adjustment software PACTware.
8.4 Instrument repair
If it is necessary to repair VEGAPULS 68 please
proceed as follows:
You can download a return form (23 KB) from our
Internet homepage www.vega.com under "Services >
Downloads > Forms and Certicates > Repair form.
Setup
VEGAPULS 68 - 4 ... 20 mA/HART 49
Maintenance and fault rectication