Process conditions
Vessel pressure horn antenna
- without swivelling holder -1 ... 40 bar (-100 ... 4000 kPaor-14.5 ...
580)
- with swivelling holder -1 ... 1 bar (-100 ... 100 kPaor-14.5 ...
14.5 psi) not sealing
vessel pressure parabolic antenna
- without swivelling holder -1 ... 6 bar (-100 ... 4000 kPaor-14.5 ... 87
psi)
- with swivelling holder -1 ... 1 bar (-100 ... 100 kPaor-14.5 ...
14.5 psi) not sealing
Process temperature (
measured on the process tting) depending on the seal of
the antenna system
- Viton -40 ... +130°C (-40 ... +266°F)
- Viton with temperature adapter -40 ... +200°C (-40 ... +392°F)
- Kalrez 2035, 6230, 6623 -15 ... +130°C (-5 ... +266°F)
- Kalrez 2035, 6230 with tempe-
rature adapter
-15 ... +200°C (-5 ... +392°F)
- Kalrez 6375 -20 ... +130°C (-
4 ... +266°F)
- Kalrez 6375 with temperature
adapter
-20 ... +200°C (-4 ... +392°F)
Vibration resistance mechanical vibrations with 4 g and 5 ...
100 Hz
4)
Data on rinsing air connection
Pressure max. 6 bar (87 psi)
Thread GA
Check valve (option, attached with Ex version)
- Material 316Ti (stainless steel 1.4571)
- Seal Viton
- for tube diameter 6 mm
- opening pressure 0.5 bar (7.3 psi)
- nominal pressure stage PN 250
4)
Tested acc. to the regulations of German Lloyd, GL directive 2
VEGAPULS 68 - 4 ... 20 mA/HART 55