User's Manual

Relay data
- AC 5 A, 253 V AC
- DC 4 A, 30 V DC; 0.2 A, 125 V DC
Breaking capacity If inductive loads or stronger currents are switched
through, the gold plating on the relay contact
surface will be permanently damaged. The contact
is then no longer suitable for switching low-level
signal circuits.
Contact material (relay contacts) AgNi or AgSnO and Au plated
Modes (adjustable)
- max. Max. detection or overow/overll protection
- min. Min. detection or dry run protection
Measuring accuracy
Hysteresis approximately 1 dB
Switching delay adjustable from 0.1 20 s (on/o)
Min. distance 100 mm (3.94 in)
Ambient conditions
Ambient, storage and transport temperature -40 +80 °C (-40 +176 °F)
Process conditions
Measured variable Limit level of bulk solids and liquids
Process pressure
- VEGAMIP 61 antenna with PTFE cover -1 4 bar/-100 400 kPa (-14.5 58 psig)
- VEGAMIP 61 antenna with PP cover -1 2 bar/-100 200 kPa (-14.5 29 psig)
- VEGAMIP 61 with mounting adapter unpressurized (IP 67)
Process temperature (thread or ange tem-
perature)
- VEGAMIP 61 antenna with PTFE cover -40 +80 °C (-40 +176 °F)
- VEGAMIP 61 antenna with PP cover -40 +80 °C (-40 +176 °F)
- VEGAMIP 61 with mounting adapter
150 mm (optional)
-40 +250 °C (-40 +482 °F)
- VEGAMIP 61 with mounting adapter
300 mm (optional)
-40 +450 °C (-40 +842 °F)
Electromechanical data
Cable entry/plug (dependent on the version)
- Single chamber housing l 1 x cable entry M20 x 1.5 (cable: ø 5 9 mm),
1 x blind stopper M20 x 1.5; attached 1 x cable
entry M20 x 1.5
or:
32 VEGAMIP R61 - Relay
9 Supplement
35786-EN-100531