6 Set up with the display and adjustment module
VEGAPULS 62 • 4 … 20 mA/HART - two-wire
36503-EN-161221
Menu Menu item Default value
Setup Measurement
loop name
Sensor
Medium Liquid/Water
Bulk solids/Crushed stones, gravel
Application Storage tank
Silo
Vessel form Vessel bottom, dished form
Vessel top, dished form
Vessel height/
Measuring range
Recommended measuring range, see
"Technical data" in the supplement
Min. adjustment Recommended measuring range, see
"Technical data" in the supplement
Max. adjustment 0,000 m(d)
Damping 0.0 s
Current output
mode
4 … 20 mA, < 3.6 mA
Current output
Min./Max.
Min. current 3.8 mA, max. current
20.5 mA
Lock adjustment Released
Display Language Like order
Displayed value Distance
Display unit m
Scalingsize Volume
l
Scaling 0.00 lin %, 0 l
100.00 lin %, 100 l
Backlight Switched on
Additional adjust-
ments
Distance unit m
Temperature unit °C
Probe length Length of the standpipe Ex factory
Linearisation
curve
Linear
HART mode Standard
Address 0
ThesensoroerstheHARTmodesstandardandMultidrop.Inthis
menu item you specify the HART mode and enter the address for
Multidrop.
Additional adjustments -
HART mode