User's Manual

82
11 Supplement
VEGAPULS 62 • 4 … 20 mA/HART - two-wire
36503-EN-161221
Pressure Withoutreuxvalve Withreuxvalve
0.6 bar (8.70 psig) 3.5 m
3
/h 1.4 m
3
/h
0.7 bar (10.15 psig) 3.7 m
3
/h 1.7 m
3
/h
0.8 bar (11.60 psig) 3.9 m
3
/h 1.8 m
3
/h
0.9 bar (13.05 psig) 4.0 m
3
/h 2.1 m
3
/h
1 bar (14.5 psig) 4.2 m
3
/h 2.2 m
3
/h
1.5 bar (21.76 psig) 5.0 m
3
/h 3.2 m
3
/h
2 bar (29.0 psig) 5.5 m
3
/h 4.5 m
3
/h
Air volume with parabolic antenna, depending on pressure (recommended area)
Pressure Withoutreuxvalve Withreuxvalve
0.5 bar (7.25 psig) 3.0 m
3
/h 1.2 m
3
/h
0.6 bar (8.70 psig) 3.2 m
3
/h 1.4 m
3
/h
0.7 bar (10.15 psig) 3.4 m
3
/h 1.7 m
3
/h
0.8 bar (11.60 psig) 3.5 m
3
/h 1.9 m
3
/h
0.9 bar (13.05 psig) 3.6 m
3
/h 2.0 m
3
/h
1 bar (14.5 psig) 3.8 m
3
/h 2.2 m
3
/h
1.5 bar (21.76 psig) 4.3 m
3
/h 3.5 m
3
/h
2 bar (29.0 psig) 4.8 m
3
/h 4.0 m
3
/h
Thread G⅛
Closure with
Ʋ Non-Ex Dust protection cover of PE
Ʋ Ex Threaded plug of 316Ti
Reuxvalve-unmounted(asoptionwithnon-Exversion,includedinthescopeofdeliverywithEx
version)
Ʋ Material 316Ti
Ʋ Seal FKM(SHSFPM70C3GLT),FFKM(Kalrez6375)
Ʋ for tube diameter 6 mm
Ʋ Opening pressure 0.5 bar (7.25 psig)
Ʋ Nominal pressure stage PN 250
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable entry M20 x 1.5, ½ NPT
Ʋ Cable gland M20 x 1,5; ½ NPT (cable ø see below table)
Ʋ Blind plug M20 x 1.5; ½ NPT
Ʋ Closing cap ½ NPT