4 Mounting
VEGAPULS 69 • 4 … 20 mA/HART - two-wire
47247-EN-170315
Fig.11:Mountingoftheradarsensorwithinowingmedium
Information:
The mounting socket should be as short as possible and its end
rounded. This reduces false echoes from the vessel mounting socket.
Fig. 12: Recommended socket mounting
Ifthemediumhasgoodreectiveproperties,youcanalsomountthe
VEGAPULS 69 on longer sockets. Approximate socket heights are
shown in the following illustration.
Information:
When mounting on longer sockets, we recommend carrying out a
false signal suppression (see chapter "Parameter adjustment").
Type of socket