User's Manual

Table Of Contents
69
9 Diagnosis, asset management and service
VEGAPULS 69 • 4 … 20 mA/HART - two-wire
47247-EN-170315
Error Reason Removal
4 … 20 mA signal not sta-
ble
Fluctuations of the
measured variable
Set damping appropriate to the instrument via the
display and adjustment module or PACTware/DTM
4 … 20 mA signal missing
Electrical connection
faulty
Check connection according to chapter "Connection
steps" and if necessary, correct according to chapter
"Wiring plan"
Voltage supply missing
Check cables for breaks; repair if necessary
Operating voltage too
low or load resistance
too high
Check, adapt if necessary
Current signal greater than
22 mA or less than 3.6 mA
Electronics module in the
sensor defective
Exchange the instrument or send it in for repair
The below tables show typical examples of application-related
measurement errors with bulk solids. A distinction is made between
measurement errors during:
Constant level
Filling
Emptying
The images in column "Errorpattern" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
1
2
Level
time
0
1 Real level
2 Level displayed by the sensor
Notes:
Whenever the sensor displays a constant value, the reason could
also be that the fault setting of the current output is set to "Hold
value"
If the level indication is too low, the reason could be a line resist-
ance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value
shows a too low or too
high level
Level
time
0
Min./max. adjustment not
correct
Adapt min./max. adjustment
Incorrectlinearizationcurve
Adaptlinearizationcurve
Treatment of measure-
ment errors with bulk
solids