User's Manual

Table Of Contents
37
6 Set up with the display and adjustment module
VEGAPULS 64 • 4 … 20 mA/HART - two-wire
51141-EN-160616
parameters to ensure optimum adjustment of the measurement. The
procedure is described in the following.
Here you can assign a suitable measurement loop name. Push the
"OK" key to start the editing. With the "+" key you change the sign and
with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
Capital letters from A … Z
Numbers from 0 … 9
Special characters + - / _ blanks
In this menu item you select the distance unit and the temperature
unit.
For the distance units you can choose between m, in and ft and for
the temperature units °C, °F and K.
This menu item allows you to adapt the sensor to the measuring
conditions.
Medium
The following options are available:
Application
The following options are available:
The following features form the basis of the applications:
- Storage tank:
Setup: large-volumed, upright cylindrical, spherical
Productspeed:slowllingandemptying
Setup - Measurement
loop name
Setup - Units
Setup - Application