ca. 10 mm
Fig. 9: Recommended socket mounting
If the reective properties of the medium are good, you can
mount VEGAPULS 62 on sockets higher than the antenna
length. You will nd recommended values for socket heights in
the following illustration. The socket end should be smooth and
burr-free, if possible also rounded. Then carry out a false echo
storage.
d
h
max.
d
1½"
50 mm/2"
80 mm/3"
100 mm/4"
150 mm/6"
200 mm
250 mm
300 mm
500 mm
800 mm
h
max.
Fig. 10: Deviating socket dimensions
With liquids, align the sensor directly towards the product
surface (as vertical as possible) to achieve optimum meas-
urement.
Sensor orientation
20 VEGAPULS 62 - 4 ... 20 mA/HART
Mounting