User's Manual

Table Of Contents
5 Connecting to power supply
5.1 Preparing the connection
Always observe the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltages are expected, overvoltage arresters should
be installed
Tip:
We recommend using VEGA overvoltage arresters ÜS-F-LB-I
and ÜSB 62-36G.X.
In hazardous areas you should take note of the appropriate
regulations, conformity and type approval certicates of the
sensors and power supply units.
Power supply and current signal are transmitted via the same
two-wire connection cable. The supply voltage range can dier
depending on the instrument version. The exact range is
stated in the "Technical data" in the "Supplement".
Provide a reliable separation between the supply circuit and
the mains circuits acc. to DIN VDE 0106 part 101. The VEGA
power supply units VEGATRENN 149AEx, VEGASTAB 690,
VEGADIS 371 as well as all VEGAMETs meet this require-
ment.
Bear in mind the following factors regarding supply voltage:
l the reduction of the output voltage of the power supply unit
under nominal load (with a sensor current of 20.5 mA or
22 mA in case of fault signal)
l The inuence of additional instruments in the circuit (see
load values in "Technical data").
VEGAPULS 62 is connected with standard two-wire cable
without screen. A outer cable diameter of 5 9 mm ensures
the seal eect of the cable gland. If electromagnetic
interference is expected which is above the test values of
EN 61326 for industrial areas, screened cable should be used.
For HART multidrop operation we recommend as standard
practice the use of screened cable.
On VEGAPULS 62 with cable gland ½ NPT and plastic
housing, a metal ½" threaded insert is moulded in the plastic
housing.
Note safety instructions
Take note of safety
instructions for Ex
applications
Select power supply
Select connection cable
Cable gland ½ NPT
VEGAPULS 62 - 4 ... 20 mA/HART 25
Connecting to power supply
28435-EN-060317