User guide
22
5Connectingtopowersupply
MINITRAC 31 • Foundation Fieldbus
41782-EN-131119
6. Insertthewireendsintotheroundopeningsoftheterminalsac-
cordingtothewiringplan
Information:
Solidcoresaswellasexiblecoreswithcableendsleevesare
inserteddirectlyintotheterminalopenings.Incaseofexiblecores
withoutendsleeves,presstherectangularlockopeningwithasmall
screwdriver;theterminalopeningisfreed.Whenthescrewdriveris
released, the terminal opening closes again.
7. Checktheholdofthewiresintheterminalsbylightlypullingon
them
Toloosenaline,insertasmallslottedscrewdriverrmlyintothe
rectangular locking opening according to the illustration
8. Connect the screen to the internal ground terminal, connect the
outergroundterminaltopotentialequalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Screwthehousingcoverbackon
Theelectricalconnectionishencenished.
Information:
The terminal blocks are pluggable and can be detached from the
electronics.Forthispurposeloosenthetwolaterallockingleversof
theterminalblockwithasmallscrewdriver.Whenlooseningthelock-
ing,theterminalblockisautomaticallysqueezedout.Itmustsnapin
placewhenre-inserted.