6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-131119
Youcanndtherotaryswitchontheelectronicsmoduleintheelec-
tronics and connection department (large cover).
Youcansimulatedierentvalues:
Pulse rate of the sensor
Process value
Switchingfunctionoftherelay
Information:
The simulation is automatically terminated 10 minutes after the last
pressing of a key.
Additional adjustments
In this menu item, the PIN is permanently activated/deactivated. Thus
youprotectthesensordataagainstunauthorizedaccessandunin-
tended changes. The default setting of the PIN is 0000.
This parameter is described in the operating instructions manual
"Display and adjustment module".
In this menu item you can set the actual date and time.
This parameter is described in the operating instructions manual
"Display and adjustment module".
Whenaresetiscarriedout,allsettings(withonlyafewexceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
Thefollowingresetfunctionsareavailable:
Basic settings: Resetting of the parameter adjustments to default
valuesatthetimeofshipment.Order-specicsettingsaredeleted.
Additional settings - PIN
Additional adjustments -
Date Time
Additional adjustments/
Reset