6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-131119
Menu Menu item Default value
Display Language Selected language
Displayed value Pulse rate
Display unit ct/s
With this function
Load parameter adjustment data from the sensor into the display
and adjustment module
Write parameter adjustment data from the display and adjustment
module into the sensor
This parameter is described in the operating instructions manual
"Display and adjustment module".
Info
Inthismenuyouwillndthefollowingmenuitems:
Instrumentname-showsinstrumentnameandserialnumber
Instrumentversion-showshardwareandsoftwareversionofthe
instrument
Dateofmanufacture-showscalibrationdateandthedateofthe
last change
DeviceID-showsthedeviceIDandthesensorTAG(PD_TAG)
Instrumentfeatures-showsfurtherinstrumentfeatures
These parameters are described in the operating instructions manual
"Display and adjustment module".
6.6 Parameter adjustment - X-ray alarm
X-rayradiation,forexample,causewithweldjointtestscaninuence
your measured values unnoticeably. Hence it is importnat to recog-
nizedinterferingX-rayradiation.
TheMINITRAC31recognizestheX-rayraditationandreactsfroman
individuallydenableraditiationvolume.
The level or density measuring instrument acts as Master and the
MINITRAC 31 operates as Slave.
Firstofall,denethefunctionoftheSlaveinstrumentbeforeyou
denetheMasterinstrument.TheMasterinstrumentcanthusimme-
diatelyrecognizetheconnectedSlave.
Forthispurpose,theSlaveinstrumentmustbedenedas"X-ray
alarm". Select under the menu item "Setup/Application" the function
"X-ray alarm".
Youcansettheaddress(MGC)ontheSlaveinstrumentatwill.Only
the address "0 - 0" is reserved for the Master instrument.
Set the address setting (MGC) on the Master instrument to "0 - 0".
Additional adjustments -
Copy instrument settings
Info
X-ray alarm