User guide

55
6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-131119
You have to enter the address of the Slave instrument in the list of
the Master instrument. This function is not possible in the display and
adjustmentmodule.TodothisyouneedPACTwarewiththerespective
DTM.
Themainmenuisdividedintovesectionswiththefollowingfunc-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments: Instrument unit, reset, date/time, copying
function
Info:Instrumentname,hardwareandsoftwareversion,dateofmanu-
facture, instrument features
Check if the correct language is already set for the display. If not, you
can change the language in the menu item "Display/Language".
StartwiththesetupofMINITRAC31.
In the main menu point "Setup", the individual submenu points
shouldbeselectedoneaftertheotherandprovidedwiththecorrect
parameters to ensure optimum adjustment of the measurement. The
procedureisdescribedinthefollowing.
Stickwiththenormalsequenceofthemenuitemsascloselyaspos-
sible.
Setup
In this menu item you can adjust the MINITRAC 31 to the integrated
isotope in the source container.
Forthispurpose,checkwhichisotopeisintegratedinthesource
container.Youcanndthisinformationonthetypelabelofthesource
container.
With this selection, the sensitivity of the sensor is adapted in an opti-
mumwaytotheisotope.
Main menu
Procedure
Setup - Isotope