9 Diagnostics and service
MINITRAC 31 • Foundation Fieldbus
41782-EN-131119
Code
Text mes-
sage
Cause Rectication
F036
Faulty pro-
gram memory
Errorduringsoftware
update
Repeatsoftwareupdate
Exchanging the electronics
F037
RAM hard-
wareerror
Error in RAM Restart instrument
Exchanging the electronics
F038
Slave signals
failure
Connection cable to the
Slave instrument interrupted
Instrumentnotdenedas
Slave instrument
Check the connection cable
to the Slave instrument
DeneinstrumentasSlave
F040
Hardware
error
Sensor defective Restart instrument
Exchanging the electronics
F041
Photomultipli-
er error
Error in the measured value
recording
Restart instrument
Exchanging the electronics
F052
Faultycong-
uration
Invalid parameter adjust-
ment
Carry out a reset
F066
Faulty adjust-
ment
Adjustment not yet carried
out
Error during adjustment or
whenenteringthelineariza-
tion table
Carrying out adjustment
Carryoutlinearization
F068
Count rate too
high
Faulty instrument settings Carry out a reset
F072
Limit exceed-
ed
Faulty instrument settings Carry out a reset
F073
Error real val-
ue correction
Faulty real value correction Carry out real value correc-
tion again
F080
System error
Instrument error Restart instrument
Call our service
F086
Communica-
tion error
Error in the Fieldbus com-
munication
Restart instrument
Call our service
F114
Error real time
clock
Discharge accumulator Readjust real time clock
F120
Filter time
error
Faulty or missing instrument
adjustment
Carrying out adjustment