Operating Instructions Radiation-based sensor for continuous level and interface measurement SOLITRAC 31 Foundation Fieldbus Document ID: 41781
Contents Contents 1 About this document 1.1 Function............................................................................................................................ 4 1.2 Target group...................................................................................................................... 4 1.3 Symbolism used................................................................................................................ 4 2 For your safety 2.1 Authorised personnel...............
Contents 9.5 9.6 Software update.............................................................................................................. 59 How to proceed in case of repair..................................................................................... 59 10 Dismounting 10.1 Dismounting steps.......................................................................................................... 60 10.2 Disposal...............................................................................
1 About this document 1 About this document 1.1 Function This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. 1.2 Target group This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 For your safety 2.1 Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use SOLITRAC 31 is a sensor for continuous level measurement. You can find detailed information on the application range in chapter "Product description".
2 For your safety 2.5 CE conformity The device fulfills the legal requirements of the applicable EC guidelines. By affixing the CE marking, VEGA confirms successful testing of the product. Only with class A instruments: The device is a class A instrument designed for use in an industrial environment. When used in a different environment, e.g., in a living area, the electromagnetic compatibility must be ensured by the user.
3 Product description 3 Product description Type plate 3.1 Configuration The nameplate contains the most important data for identification and use of the instrument: 1 2 3 4 5 6 10 9 9 7 8 Fig.
3 Product description • Electronics versions Modification status, electronics as of -01 The instrument is available in different electronics versions. Each version can be identified via the product code on the type label: • Standard electronics type PT30E-XX Scope of delivery The scope of delivery encompasses: Application area The instrument is suitable for applications in liquids and bulk solids in vessels under difficult process conditions.
3 Product description Transport inspection The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with. Storage Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
3 Product description The handling of radioactive substances is regulated by law. The radiation protection rules of the country in which the system is operated apply first and foremost. In Germany, for example, the current radiation protection ordinance (StrlSchV) based on the Atomic Energy Law (AtG) applies. The following points are important for measurement with radiationbased methods: Handling permit A handling permit is required for operation of a system using gamma rays.
3 Product description for control areas in the guideline of the respective authority (in Germany, for example, the radiation protection ordinance). 41781-EN-131119 We are at your disposal for further information concerning radiation protection and regulations in other countries.
4 Mounting 4 Mounting Switch off source 4.1 General instructions The source container is part of the measuring system. In case the source container is already equipped with an active isotope, the source container must be locked before mounting. Danger: Before mounting; make sure that the source is securely closed. Use a padlock to secure the source container in the closed condition and prevent it from being inadvertently opened.
4 Mounting Installation position 4.2 Mounting instructions Note: Along with the planning, our specialists will analyse the conditions of the measurement loop to dimension the radiation source (isotope) accordingly. You get a "Source Sizing" document specifying the required source activity and containing all relevant mounting information for your measuring point. You must follow the instructions of this "Source Sizing" document in addition to the following mounting instructions.
4 Mounting Summation To measure the level in very high vessels, multiple instruments can be cascaded. Cascading means that two or several instruments are connected which can together cover a longer measuring range. You can find the exact number of Slaves in the Safety Manual. L2 h L1 1 Fig.
4 Mounting Vessel with heat insulation On vessels with temperature insulation, the sensor and the source container should be preferably mounted outside of the tank insulation. Fig. 5: Level measurement on a heated vessel with heat insulation If this is not possible, make sure that there is a sufficiently large cut-out in the tank insulation for mounting the sensor and the source container. Make sure that the max. ambient temperature of the sensor is not exceeded. Protection against heat If the max.
5 Connecting to power supply 5 Connecting to power supply Safety instructions 5.1 Preparing the connection Always keep in mind the following safety instructions: • • Connect only in the complete absence of line voltage If overvoltage surges are expected, overvoltage arresters should be installed Voltage supply Power supply and digital bus cable via separate two-wire connection cables. Power is supplied via the H1 voltage supply.
5 Connecting to power supply T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). Low-frequency potential equalisation currents are thus suppressed, but the protective effect against high frequency interference signals remains.
5 Connecting to power supply 6. Insert the wire ends into the round openings of the terminals according to the wiring plan Information: Solid cores as well as flexible cores with cable end sleeves are inserted directly into the terminal openings. In case of flexible cores without end sleeves, press the rectangular lock opening with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the terminal opening closes again. 7.
5 Connecting to power supply 5.2 Connection - Level measurement 10 /N 2 PE 4 5 2 6 9 12 11 10 11 3 9 22 21 20 19 18 17 16 15 14 13 12 1 1 /L 22 1 Electronics and connection compartment - Non-Ex instruments and instruments with nonintrinsically safe current output Non-Ex instruments and instruments with non-intrinsically safe current output 8 7 5 6 4 Fig.
5 Connecting to power supply 9 8 /N 2 PE 4 5 2 6 12 11 22 21 20 19 18 17 16 15 14 13 12 /L 1 1 22 1 Electronics and connection compartment - Instruments with intrinsically safe current output instructions are part of the scope of delivery and come with the Exapproved instruments. 7 6 4 5 3 Fig.
5 Connecting to power supply 5.3 Connection - Level detection 10 /N 2 PE 4 5 2 6 9 12 11 10 11 3 9 22 21 20 19 18 17 16 15 14 13 12 1 1 /L 22 1 Electronics and connection compartment - Non-Ex instruments and instruments with nonintrinsically safe current output Non-Ex instruments and instruments with non-intrinsically safe current output 8 7 5 6 4 Fig.
5 Connecting to power supply 9 8 /N 2 PE 4 5 2 6 12 11 22 21 20 19 18 17 16 15 14 13 12 /L 1 1 22 1 Electronics and connection compartment - Instruments with intrinsically safe current output instructions are part of the scope of delivery and come with the Exapproved instruments. 7 6 4 5 3 Fig.
5 Connecting to power supply Electronics and connection compartment - summation 5.4 Connection - Summation To measure the level in very high vessels, multiple instruments can be cascaded. Cascading means that two or several instruments are connected which can together cover a longer measuring range. The instrument acts as Master and all other instruments operate as Slaves. The pulse rates of all instruments are summed in the Master instrument and converted into a common signal.
5 Connecting to power supply M S S 2 4 bus 2 6 4 5 9 bus bus 2 4 9 5 6 12 11 10 11 S 9 12 11 10 11 22 21 20 19 18 17 16 15 14 13 12 1 22 1 6 12 11 10 11 M 22 21 20 19 18 17 16 15 14 13 12 1 22 1 5 22 21 20 19 18 17 16 15 14 13 12 1 22 1 bus bus bus S M Master instrument S Slave instrument 24 SOLITRAC 31 • Foundation Fieldbus 41781-EN-131119 Fig. 15: Electronics and connection compartment with cascading of several instruments.
5 Connecting to power supply Information: For example, a radial connection would be also possible as an alternative. Take note of the polarity. 41781-EN-131119 The selection of the two terminal pairs is individual.
6 Set up with the display and adjustment module 6 Set up with the display and adjustment module Mount/Dismount display and adjustment module 6.1 Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed again at any time. It is not necessary to interrupt the power supply. Proceed as follows: 1. Unscrew the small housing cover 2.
6 Set up with the display and adjustment module 6.2 Adjustment system 1 2 Fig.
6 Set up with the display and adjustment module Information: In this operating instructions manual, the instrument-specific parameters are described. Further general parameters are described in the operating instructions manual "Display and adjustment module". Instrument start Caution: During the first setup or after an instrument reset the instrument starts with preset standard values. These value are not suitable for your application and must be replaced by real values.
6 Set up with the display and adjustment module Through this selection, the sensitivity of the sensor is adapted perfectly to the isotope. The normal reduction of the emitter activity is hence considered through the radioactive decay. The SOLITRAC 31 requires this information of the automatic decay compensation. This ensures an interference-free measurement over the complete life time of the gamma emitter - an annual recalibration is not necessary.
6 Set up with the display and adjustment module These settings influence the current output of the sensor. Enter in the menu window "Max. process value" the max. level (full), for example in "m". This corresponds to an output current of 20 mA. Enter in the menu window "Min. process value" the min. level (empty), for example in "m". This corresponds to an output current of 4 mA. Setup - Linearization In this menu item you can carry out the adjustment of the sensor.
6 Set up with the display and adjustment module Select "Linearization - New" to enter the first point. Select "Determine count rate" to enter the first point. The determination of the actual count rate lasts 2 minutes. After the count rate has been determined, you can accept the value (ct/s). The count rate is stated in ct/s. This is the number of counts per second, i.e. the measured radioactive radiation dose actually reaching the sensor. Enter now the corresponding level (m).
6 Set up with the display and adjustment module Because you have already set the background radiation to 0 in the menu item: "Setup/Background radiation", the measured pulse rate must be around 0 ct/s. Show diagram This menu item is only available if a linearization was already carried out. Show table In this menu item you can show the individual value pairs of the linearization. Linearization - Delete You can also delete individual linearization points.
6 Set up with the display and adjustment module Setup - Real value correction If you know the level at a certain height, you can enter in this menu item the determined real level to correct the measured value. The function shifts the linearization curve to this determined point. With this procedure the measurement can be adapted exactly to the conditions in the vessel. Setup - Relay In this menu item you can activate the relay output and determine its function as well as the switching points.
6 Set up with the display and adjustment module Caution: Independent of the selected reference value, the relay will deenergize in case of failure. Lock setup - adjustment With this menu item you safeguard the sensor parameters against unauthorized or unintentional modifications. This menu item is described in the operating instructions manual "Display and adjustment module". Display Display - Language With this parameter you can change the display language.
6 Set up with the display and adjustment module Diagnosis - Simulation In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g. through downstream indicating instruments or the input card of the control system. Note: To carry out a simulation with the display and adjustment module, you have to switch on the simulation switch on the electronics module (switch position 1).
6 Set up with the display and adjustment module The following reset functions are available: Basic settings: Resetting of the parameter adjustments to default values at the time of shipment. Order-specific settings are deleted. Default settings: Resetting of the parameter adjustment like under "Basic settings". In addition, special parameters are reset to default values. Order-specific settings are deleted.
6 Set up with the display and adjustment module Menu Menu item Default value Setup Measurement loop name Sensor Isotope Cs-137 Application Level Background radiation 0 ct/s Unit of the process value % Temperature unit ° C Adjustment min. Min. process value = 0 % Adjustment max. Max.
6 Set up with the display and adjustment module Info Info In this menu you will find the following menu items: • • • • • Instrument name - shows instrument name and serial number Instrument version - shows hardware and software version of the instrument Date of manufacture - shows calibration date and the date of the last change Device ID - shows the device ID and the sensor TAG (PD_TAG) Instrument features - shows further instrument features These parameters are described in the operating instruc
6 Set up with the display and adjustment module Setup: Settings, e.g.
6 Set up with the display and adjustment module This menu item enables adaptation of the sensor to the requested application. You can choose between the following applications: "Level", "Point level" or "Summation Slave". Setup - Outputs In this menu item you can activate the function of the current output. When the output is activated, the instrument remains in its function as a Slave, but the 4 … 20 mA output of the SOLITRAC 31 can be also used als single instrument.
6 Set up with the display and adjustment module Diagnosis: Information, for example, of device status, peak value, simulation Additional adjustments: Instrument unit, reset, date/time, copying function Info: Instrument name, hardware and software version, date of manufacture, instrument features Procedure Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language". Start with the setup of SOLITRAC 31.
6 Set up with the display and adjustment module Setup - Background radiation The natural radiation on earth influences the accuracy of the measurement. With this menu item the natural background radiation can be faded out. For this purpose, the SOLITRAC 31 measures the natural background radiation and sets the pulse rate to zero. In the future, the pulse rate from this background radiation will be automatically deducted from the total pulse rate.
6 Set up with the display and adjustment module Empty the vessel until the sensor is uncovered. For this enter the requested pulse rate manually or let the rate be determined by SOLITRAC 31. Automatic determination of the pulse rate should be given preference. The pulse rate is entered in ct/s. This is the number of counts per second, i.e. the measured gamma radiation reaching the sensor.
6 Set up with the display and adjustment module You thus get the min. pulse rate (ct/s) for the "covered" adjustment. Enter the requested pulse rate manually or let the rate be determined by SOLITRAC 31. Automatic determination of the pulse rate should be given preference. You can enter the adjustment point (ct/s) manually. You can let the adjustment point be determined by SOLITRAC 31.
6 Set up with the display and adjustment module Overfill protection = the relay will deenergise (safe condition) when the max. level is reached. Dry run protection = the relay will deenergise (safe condition) when the min. level is reached. Make sure that you have selected the correct characteristics. See menu item "Setup/Current output mode". Lock setup - adjustment With this menu item you safeguard the sensor parameters against unauthorized or unintentional modifications.
6 Set up with the display and adjustment module The greater the difference in the pulse rate between covered and uncovered status, the greater the differential value (Delta I) and the more reliable the measurement. The automatically calculated damping is also oriented around the Delta I value. The higher the value, the lower the damping. A Delta I value below 10 % is an indication for a critical measurement.
6 Set up with the display and adjustment module Additional adjustments Additional settings - PIN In this menu item, the PIN is permanently activated/deactivated. Thus you protect the sensor data against unauthorized access and unintended changes. The default setting of the PIN is 0000. This parameter is described in the operating instructions manual "Display and adjustment module". Additional adjustments Date Time In this menu item you can set the actual date and time.
6 Set up with the display and adjustment module Menu Menu item Default value Setup Measurement loop name Sensor Isotope Cs-137 Application Limit level Adjustment mode Single point adjustment Adjustment - uncovered 90000 ct/s Adjustment - covered 9000 ct/s Delta I 90 % Background radiation 0 ct/s Temperature unit ° C Adjustment mode 1-point Adjustment uncovered 900000 ct/s Delta I 90 % Mode relay Overfill protection Block adjustment Released Language Selected language Displa
6 Set up with the display and adjustment module • • • Date of manufacture - shows calibration date and the date of the last change Device ID - shows the device ID and the sensor TAG (PD_TAG) Instrument features - shows further instrument features These parameters are described in the operating instructions manual "Display and adjustment module". 6.6 Saving the parameter adjustment data We recommended noting the adjusted data, e.g.
7 Setup with PACTware 7 Setup with PACTware Via the interface adapter directly on the sensor 7.1 Connect the PC 2 1 3 Fig. 18: Connection of the PC directly to the sensor via the interface adapter 1 USB cable to the PC 2 Interface adapter VEGACONNECT 4 3 ">Sensor Information: The interface adapter VEGACONNECT 3 is not suitable for connection to the sensor. Prerequisites 7.
7 Setup with PACTware Fig. 19: Example of a DTM view Standard/Full version All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project configuration simplifies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
8 Set up with other systems 8 Set up with other systems 8.1 DD adjustment programs Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM. The files can be downloaded at www.vega.com/downloads under "Software". 8.2 Field Communicator 375, 475 Device descriptions for the instrument are available as EDD for parameter adjustment with the Field Communicator 375 or 475.
9 Diagnostics and service 9 Diagnostics and service 9.1 Maintenance If the device is used correctly, no maintenance is required in normal operation. The corresponding source container must be checked in regular intervals. You can find further information in the operating instructions manual of the source container. 9.2 Status messages The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650.
9 Diagnostics and service Maintenance: Due to external influences, the instrument function is limited. The measurement is affected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup). This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
9 Diagnostics and service Code Cause Rectification F036 –– Error during software update –– Repeat software update –– Exchanging the electronics F037 –– Error in RAM –– Restart instrument –– Exchanging the electronics F038 Slave signals failure –– Connection cable to the –– Check the connection cable Slave instrument interrupted to the Slave instrument –– Instrument not defined as –– Define instrument as Slave Slave instrument F040 –– Sensor defective F041 –– Error in the measured value –– Res
9 Diagnostics and service Code Cause F121 –– Slave instruments not found –– Check Slave instruments –– Check Slave list in Master instrument –– Slave instrument with wrong address F122 –– Instrument addresse was assigned several times –– Change instrument addresses F123 X-ray alarm –– External instruments cause radiation –– Determine reason for X-ray alarm –– In case of brief X-ray radiation: Monitor switching outputs for this time manually F124 –– Radiation dose too high –– Determine reason fo
9 Diagnostics and service Code Cause Rectification S017 –– Accuracy outside the specification –– Correct adjustment data S025 –– Bad linearization table –– Carry out linearization S038 –– Slave outside the specifica- –– Check Slave tion S125 –– Ambient temperature too high/too low Text message Accuracy outside the specification Bad linearization table Slave outside the specification Ambient temperature too high/too low –– Protect instrument with isolating material against extreme temperatures
9 Diagnostics and service Check output signal (level detection) The following table describes possible faults that may not generate an error message: Error Cause Rectification The instrument signals covered without covering with the medium Voltage supply missing Check cables for breaks; repair if necessary Operating voltage too low Check, adapt if necessary Electrical connection faulty Check connection according to chapter "Connection steps" and if necessary, correct according to chapter "Wiring
9 Diagnostics and service 9.5 Software update The following components are required to update the sensor software: • • • • • Sensor Voltage supply Interface adapter VEGACONNECT PC with PACTware Current sensor software as file You can find the actual sensor software as well as detailed information of the procedure under "www.vega.com/downloads" and "Software". You can find information about the installation in the download file.
10 Dismounting 10 Dismounting 10.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 10.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
11 Supplement 11 Supplement 11.1 Technical data General data 316L corresponds to 1.4404 or 1.
1 2 L 11 Supplement Fig. 21: Data of the input variable 1 2 L Min. level (red marking line) Max. level Measuring range Measuring range (L) Analogue input ƲƲ Input type ƲƲ Internal load ƲƲ Type of input - Open Collector ƲƲ Type of input - Relay contact 62 4 … 20 mA, passive 250 Ω 41781-EN-131119 Switching input 500 … 3000 mm (19.69 … 118.
11 Supplement Output variable - Level measurement Output ƲƲ Signal ƲƲ Physical layer Damping (63 % of the input variable) Channel Numbers according to IEC 61158-2 0 … 999 s, adjustable ƲƲ Channel 1 Process value (level) ƲƲ Channel 9 Pulse rate ƲƲ Channel 8 Transmission rate Current value Resolution, digital Output variable - Level detection Output ƲƲ Signal ƲƲ Physical layer Damping (63 % of the input variable) Channel Numbers Electronics temperature 31.25 Kbit/s 10 mA, ±0.5 mA > 0.1 mm (0.
11 Supplement Transistor output Output Floating transistor output, permanently shortcircuit-proof Voltage loss < 1 V Load current Switching voltage Blocking current < 400 mA < 55 V DC < 10 µA Accuracy (according to DIN EN 60770-1) Process reference conditions according to DIN EN 61298-1 ƲƲ Temperature +18 … +30 °C (+64 … +86 °F) ƲƲ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.
11 Supplement Spring-loaded terminals for wire cross-section ƲƲ Massive wire, cord ƲƲ Stranded wire with end sleeve Display and adjustment module Display element Measured value indication ƲƲ Number of digits ƲƲ Size of digits Adjustment elements Protection rating ƲƲ unassembled ƲƲ mounted into the housing without cover Materials ƲƲ Housing ƲƲ Inspection window 0.2 … 2.5 mm² (AWG 24 … 14) 0.2 … 1.
11 Supplement For that reason the associated approval documents of these instruments must be carefully noted. They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals". 11.2 Supplementary information Foundation Fieldbus The following table gives you an overview of the instrument versions and the corresponding device descriptions, the electrical characteristics of the bus system as well as the applied function blocks.
11 Supplement Fig. 22: Schematic presentation function block Analog Input (AI) Parameter list 41781-EN-131119 The following table gives you an overview of the parameters used. FF descriptor Rel. In- Description dex Unit PRIMARY_VALUE 13 PRIMARY_VALUE (Linearized value). This is the process value after min/max adjustment and Linearization with the status of the transducer block.
11 Supplement Rel. In- Description dex Unit FF_ELECTRONIC_ TEMPERATURE 27 Electronics temperature FF_ELECTRONIC_ TEMPERATURE_UNIT FF_ELECTRONIC_ TEMPERATURE_UNIT 28 Selected unit code for “FF_ELECTRONIC_TEMPERATURE” - FF_COUNT_RATE 29 Count rate FF_COUNT_RATE_ UNIT FF_COUNT_RATE_ UNIT 30 Selected unit code for “FF_COUNT_RATE” - DEVICE_TAG 31 Tagname - DEVICE_NAME 32 Device type - DEVICE_STATE 33 Error code - PEAK_MEAS_VAL_MIN 34 Pulse rate (min.
11 Supplement FF descriptor Rel.
11 Supplement Priority Description User Error Codes NE-107 Status 29 Software failure 80 FAILURE 28 Paramererization corrupt 72 FAILURE 27 Undefined 27 - - 26 Undefined 26 - - 25 Paramererization error 16, 17, 25, 52, 53, 57, 66, 117, 120 FAILURE 24 Conflict in MGC 121, 122, 141 FAILURE 23 Communication error in MGC 8 FAILURE 22 MGC slave reports error 38 FAILURE 21 Undefined 21 - - 20 Undefined 20 - - 19 Undefined 19 - - 18 Detector temperature critical 125
11 Supplement 9 mm (0.35") Aluminium and stainless steel housing 119 mm (4.69") M20x1,5/ ½ NPT 169 mm (6.65") 116,5 mm (4.59") 242,5 mm (9.55") 54,5 mm (2.15") 9 mm (0.35") 175 mm (6.89") 90 mm (3.54") 143,5 mm (5.65") 100 mm (3.94") 41781-EN-131119 Fig.
11 Supplement 22 mm (0.87") 242,5 mm (9.55") SOLITRAC 31 36 mm (1.42") L ø 76,2 mm (3") Fig.
11 Supplement SOLITRAC 31 - Mounting example 220 mm (8.66") 101,6 mm (4.00") 70 mm (2.76") 195 mm (7.68") L 152,5 mm (6.00") 54 mm (2.13") 51 mm (2.01") 14 mm (0.55") 31,8 mm (1.5") 38 mm (1.50") 183 mm (7.20") 38 mm (1.50") 127 mm (5.00") 14,3 mm (0.56") Fig.
11 Supplement 11.4 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. Only in U.S.A.: Further information see patent label at the sensor housing. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.
INDEX INDEX ––Co-60 28, 39, 41 ––Cs-137 28, 39, 41 A Adjustment 29 ––System 27 Adjustment data 34, 45 Adjustment mode 42 Adjustment point 42 Application 29, 39, 41 B Background radiation 29, 42 Block adjustment 34, 40, 45 C Cable entry 16 Cascading 38 Check signal 57, 58 Connection procedure 17 Connection technology 17 Control areas 10 D Damping 32, 46 Date 35, 47 Date of manufacture 38, 48 Default values 36, 47 Delta I 43 Device name 38, 48 Device settings Copy 37, 48 Displayed value 34, 45 E ED
INDEX V Voltage supply 65 W Water cooling 15 41781-EN-131119 76 SOLITRAC 31 • Foundation Fieldbus
41781-EN-131119 Notes SOLITRAC 31 • Foundation Fieldbus 77
Notes 41781-EN-131119 78 SOLITRAC 31 • Foundation Fieldbus
41781-EN-131119 Notes SOLITRAC 31 • Foundation Fieldbus 79
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2013 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.