Operating Instructions Pressure transmitter with MINI-CERTEC® measuring cell VEGABAR 54 4 … 20 mA Document ID: 36725
Contents Contents About this document 1.1 Function............................................................................................................................ 4 1.2 Target group...................................................................................................................... 4 1.3 Symbolism used................................................................................................................ 4 2 For your safety 2.1 Authorised personnel....................
Contents 8 Maintenance and fault rectification 8.1 Maintain.......................................................................................................................... 48 8.2 Rectify faults.................................................................................................................... 48 8.3 Calculation of total deviation (according to DIN 16086)................................................... 49 8.4 Exchanging the electronics module....................................
1 About this document 1 About this document 1.1 Function 1.2 Target group 1.3 Symbolism used This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 2.1 For your safety Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGABAR 54 is a pressure transmitter for measurement of gauge pressure, absolute pressure and vacuum.
2 For your safety 2.6 CE conformity 2.7 Measuring range - permissible process pressure This device fulfills the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a confirmation of successful testing. You can find the CE conformity declaration in the download area of "www.vega.com". Due to the application, a measuring cell with a measuring range higher than the permissible pressure range of the process fitting may have been integrated.
3 Product description 3 Product description 3.
3 Product description 1 14 2 13 3 4 5 6 7 12 8 11 9 10 Fig.
3 Product description The CERTEC® measuring cell is also equipped with a temperature sensor. The temperature value can be displayed via the display and adjustment module as well as processed via the signal output (with digital versions). Seal concept As a standard feature, the CERTEC® measuring cell is equipped with a lateral, recessed seal. Instruments with double seal have an additional front seal. Instruments with hygienic fitting are equipped with a gap-free form seal.
3 Product description Storage Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
4 Mounting 4 4.1 Mounting General instructions Suitability for the process Make sure that all parts of the instrument coming in direct contact conditions with the process, especially the sensor element, process seal and process fitting, are suitable for the existing process conditions, such as process pressure, process temperature as well as the chemical properties of the medium. You can find the specifications in chapter "Technical data" and on the nameplate.
4 Mounting 1 1 1 2 2 2 2 1 Fig. 4: Position of the filter element 1 2 Filter element Blind plug Caution: Due to the filter effect, the pressure compensation is time delayed. When opening/closing the housing cover quickly, the measured value can change for a period of approx. 5 s by up to 15 mbar. Information: Make sure that the filter element is always free of buildup during operation. A high-pressure cleaner may not be used for cleaning.
4 Mounting 4.2 Mounting steps Welding the socket For mounting VEGABAR 54, a welded socket is required. You can find these components in the supplementary instructions manual "Welded socket and seals". Sealing/Screwing in threaded versions Use the seal fitting belonging to the instrument, or in case of NPT connections, a high-resistance sealing material. → Screw VEGABAR 54 into the welded socket. Tighten the hexagon on the process fitting with a suitable wrench. Wrench size, see chapter "Dimensions".
4 Mounting Tip: Mount the wall mounting plate so that the cable entry of the socket housing points downward. The socket housing can be displaced by 180° to the wall mounting plate. Warning: The four screws of the socket housing must only be hand screwed. A torque > 5 Nm (3.688 lbf ft) can damage the wall mounting plate.
5 Connecting to power supply 5 Note safety instructions Connecting to power supply 5.1 Preparing the connection Always keep in mind the following safety instructions: • • Connect only in the complete absence of line voltage If overvoltage surges are expected, overvoltage arresters should be installed Tip: We recommend using VEGA overvoltage arresters B63-48 and ÜSB 62-36G.X.
5 Connecting to power supply Cable screening and grounding If screened cable is necessary, connect the cable screen on both ends to ground potential. In the sensor, the screen must be connected directly to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisation (low impedance). If potential equalisation currents are expected, the connection on the processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V).
5 Connecting to power supply 9. Check the hold of the wires in the terminals by lightly pulling on them 10. Connect the screen to the internal ground terminal, connect the outer ground terminal to potential equalisation 11. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable 12. Screw the housing cover back on The electrical connection is hence finished. Fig. 7: Connection steps 6 and 7 5.
5 Connecting to power supply Electronics and connection compartment 4...20mA 1 1 Display 2 2 Fig. 8: Electronics and connection compartment, single chamber housing 1 2 Spring-loaded terminals for voltage supply Ground terminal for connection of the cable screen Wiring plan 4...20mA 1 Display 2 1 Fig. 9: Wiring plan, single chamber housing 1 Wire assignment, connection cable Voltage supply/Signal output 5.4 Wiring plan - version IP 66/IP 68, 1 bar 1 2 Fig.
5 Connecting to power supply 5.5 Overview Wiring plan, external housing with version IP 68 Fig. 11: VEGABAR 54 in IP 68 version 25 bar and axial cable outlet, external housing Electronics and connection compartment for power supply 4...20mA 1 1 Display 2 2 3 Fig.
5 Connecting to power supply Terminal compartment, housing socket 1 2 3 4 1 2 5 3 4 6 Fig. 13: Connection of the sensor in the housing socket 1 2 3 4 5 6 Brown Blue Yellow White Shielding Breather capillaries Wiring plan external electronics 4...20mA 1 Display 2 1 Fig. 14: Wiring plan external electronics 1 Switch-on phase 5.6 Switch-on phase After connecting VEGABAR 54 to power supply or after a voltage recurrence, the instrument carries out a self-check for approx.
5 Connecting to power supply 36725-EN-130321 Then the corresponding current is outputted to the cable (the value corresponds to the actual level as well as the settings already carried out, e.g. factory setting).
6 Set up with the display and adjustment module PLICSCOM 6 Set up with the display and adjustment module PLICSCOM 6.1 Short description The display and adjustment module is used for measured value display, adjustment and diagnosis.
6 Set up with the display and adjustment module PLICSCOM Fig. 15: Insert display and adjustment module Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required. 6.3 Adjustment system 2 1 1.1 3 36725-EN-130321 Fig.
6 Set up with the display and adjustment module PLICSCOM • • • –– Save value [->] key to select: –– Menu change –– Select list entry –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– interrupt input –– Jump to next higher menu Adjustment system The sensor is adjusted via the four keys of the display and adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx.
6 Set up with the display and adjustment module PLICSCOM The actual measured value is also displayed in the menu items for min./max. adjustment. Select unit In this menu item you select the adjustment unit as well as the unit for the temperature indication in the display. To select the adjustment unit (in the example switching over from bar to mbar), proceed as follows:1) 1. Push the [OK] button in the measured value display, the menu overview is displayed.
6 Set up with the display and adjustment module PLICSCOM Density 0001000 kg/dm³ 6. Enter the requested density value with [->] and [+], confirm with [OK] and move to position correction with [->]. The adjustment unit is thus switched over from bar to m. Proceed as follows to select the temperature unit:2) 1. Activate the selection with [OK] and select "Temperature unit with [->]. 2. Activate the selection with [OK] and select the requested unit with [->] (e.g. °F). 3. Confirm with [OK].
6 Set up with the display and adjustment module PLICSCOM If the adjustment ranges are exceeded, the message "Outside parameter limits" appears. The editing procedure can be aborted with [ESC] or the displayed limit value can be accepted with [OK]. Carry out max. adjustment Proceed as follows: 1. Edit the % value in the menu item "Max. adjustment" with [OK]. Max. adjustment +100.0 % = +1000.0 mbar 0000.
6 Set up with the display and adjustment module PLICSCOM The display and adjustment module enables the adjustment without filling or pressure. Thanks to this, you can carry out your settings already in the workshop without the instrument having to be installed. The actual measured value is displayed in addition to the menu items for zero/span adjustment. Select application "Process pressure measurement" VEGABAR 54 is preset to application "Level measurement".
6 Set up with the display and adjustment module PLICSCOM The adjustment unit is thus switched over from bar to mbar. Proceed as follows to select the temperature unit:4) 1. Activate the selection with [OK] and select "Temperature unit with [->]. 2. Activate the selection with [OK] and select the requested unit with [->] (e.g. °F). 3. Confirm with [OK]. The temperature unit is hence switched over from °C to °F. Carry out a position correction Proceed as follows: 1.
6 Set up with the display and adjustment module PLICSCOM Carry out span adjustment Proceed as follows: 1. Edit the mbar value in the menu item "span" with [OK]. Span 100.0 % = +1000.0 mbar P 0000.0 mbar Information: The displayed pressure for 100 % corresponds to the nominal measuring range of the sensor (in the above example 1 bar = 1000 mbar). 2. Set the requested mbar value with [->] and [OK]. 3. Confirm with [OK] and move to the menu overview with [ESC]. The span adjustment is finished.
6 Set up with the display and adjustment module PLICSCOM • • • • Scaling Current output Unit of measurement Language • • • • SIL HART mode5) PIN Application The following safety-relevant data are not read out or written: Copy sensor data Copy sensor data? Reset The reset function resets all parameters adjusted by the user to the delivery status and the peak values to the actual values.
6 Set up with the display and adjustment module PLICSCOM Menu section Function Reset value Service Current output - characteristics 4 … 20 mA Current output - failure < 3.6 mA Current output - min. current 3.8 mA Current output - max. current 20.
6 Set up with the display and adjustment module PLICSCOM Basic adjustment ▶ Basic adjustment Display 1 Diagnostics Service Info Unit Unit of measurement 1.1 bar ▼ Temperature unit Offset 1.2 = P 0000 mbar 1.5 Min. adjustment 000.0 % 1.6 Sensor-TAG Linear ▼ Max. adjustment 100.00 % 1.4 = 0.0 mbar 100.00 mbar 0.0 mbar Linearization curve 1s 1.3 = 0.2 mbar °C ▼ Damping Position correction 0.0 mbar 1.
6 Set up with the display and adjustment module PLICSCOM Service Basic adjustment Display 4 Diagnostics ▶ Service Info Current output Characteristic: 4-20 mA ▼ 4.1 Simulation 4.2 Start simulation ▼ Fail.mode: < 3.6 mA ▼ Min. current: 3.8 mA ▼ Reset 4.3 Select reset ▼ Language 4.4 Deutsch ▼ max. current: 20.5 mA ▼ Copy sensor data 4.7 Copy sensor data? PIN 4.8 Application 4.
7 Set up with the display and adjustment module PLICSCOM 7 Set up with the display and adjustment module PLICSCOM 7.1 Short description The display and adjustment module is used for measured value display, adjustment and diagnosis.
7 Set up with the display and adjustment module PLICSCOM Fig. 17: Insert display and adjustment module Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required. 7.3 Adjustment system 2 1 1.1 3 Fig.
7 Set up with the display and adjustment module PLICSCOM • • • –– Save value [->] key to select: –– Menu change –– Select list entry –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– interrupt input –– Jump to next higher menu Adjustment system The sensor is adjusted via the four keys of the display and adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx.
7 Set up with the display and adjustment module PLICSCOM The actual measured value is also displayed in the menu items for min./max. adjustment. Select unit In this menu item you select the adjustment unit as well as the unit for the temperature indication in the display. To select the adjustment unit (in the example switching over from bar to mbar), proceed as follows:6) 1. Push the [OK] button in the measured value display, the menu overview is displayed.
7 Set up with the display and adjustment module PLICSCOM Density 0001000 kg/dm³ 6. Enter the requested density value with [->] and [+], confirm with [OK] and move to position correction with [->]. The adjustment unit is thus switched over from bar to m. Proceed as follows to select the temperature unit:7) 1. Activate the selection with [OK] and select "Temperature unit with [->]. 2. Activate the selection with [OK] and select the requested unit with [->] (e.g. °F). 3. Confirm with [OK].
7 Set up with the display and adjustment module PLICSCOM If the adjustment ranges are exceeded, the message "Outside parameter limits" appears. The editing procedure can be aborted with [ESC] or the displayed limit value can be accepted with [OK]. Carry out max. adjustment Proceed as follows: 1. Edit the % value in the menu item "Max. adjustment" with [OK]. Max. adjustment +100.0 % = +1000.0 mbar 0000.
7 Set up with the display and adjustment module PLICSCOM The display and adjustment module enables the adjustment without filling or pressure. Thanks to this, you can carry out your settings already in the workshop without the instrument having to be installed. The actual measured value is displayed in addition to the menu items for zero/span adjustment. Select application "Process pressure measurement" VEGABAR 54 is preset to application "Level measurement".
7 Set up with the display and adjustment module PLICSCOM The adjustment unit is thus switched over from bar to mbar. Proceed as follows to select the temperature unit:9) 1. Activate the selection with [OK] and select "Temperature unit with [->]. 2. Activate the selection with [OK] and select the requested unit with [->] (e.g. °F). 3. Confirm with [OK]. The temperature unit is hence switched over from °C to °F. Carry out a position correction Proceed as follows: 1.
7 Set up with the display and adjustment module PLICSCOM Carry out span adjustment Proceed as follows: 1. Edit the mbar value in the menu item "span" with [OK]. Span 100.0 % = +1000.0 mbar P 0000.0 mbar Information: The displayed pressure for 100 % corresponds to the nominal measuring range of the sensor (in the above example 1 bar = 1000 mbar). 2. Set the requested mbar value with [->] and [OK]. 3. Confirm with [OK] and move to the menu overview with [ESC]. The span adjustment is finished.
7 Set up with the display and adjustment module PLICSCOM The following data are read out or written with this function: • • • • • • • • • • • Measured value presentation Adjustment Damping Linearization curve Sensor-TAG Displayed value Display unit Scaling Current output Unit of measurement Language • • • • SIL HART mode10) PIN Application The following safety-relevant data are not read out or written: Copy sensor data Copy sensor data? Reset The reset function resets all parameters adjusted by th
7 Set up with the display and adjustment module PLICSCOM Menu section Function Reset value Display Displayed value 1 bar Displayed value 2 % Display unit Volume/l Scaling 0.00 to 100.0 Service Decimal point indication 8888.8 Current output - characteristics 4 … 20 mA Current output - failure < 3.6 mA Current output - min. current 3.8 mA Current output - max. current 20.
7 Set up with the display and adjustment module PLICSCOM Basic adjustment ▶ Basic adjustment Display 1 Diagnostics Service Info Unit Unit of measurement 1.1 bar ▼ Temperature unit Offset 1.2 = P 0000 mbar 1.5 Min. adjustment 000.0 % 1.6 Sensor-TAG Linear ▼ Max. adjustment 100.00 % 1.4 = 0.0 mbar 100.00 mbar 0.0 mbar Linearization curve 1s 1.3 = 0.2 mbar °C ▼ Damping Position correction 0.0 mbar 1.
7 Set up with the display and adjustment module PLICSCOM Service Basic adjustment Display 4 Diagnostics ▶ Service Info Current output Characteristic: 4-20 mA ▼ 4.1 Simulation 4.2 Start simulation ▼ Fail.mode: < 3.6 mA ▼ Min. current: 3.8 mA ▼ Reset 4.3 Select reset ▼ Language 4.4 Deutsch ▼ max. current: 20.5 mA ▼ Copy sensor data 4.7 Copy sensor data? PIN 4.8 Application 4.9 Level ▼ Enable? Info Basic adjustment Display 5 Diagnostics Service ▶ Info Instrument type 5.
8 Maintenance and fault rectification 8 Maintenance Maintenance and fault rectification 8.1 Maintain If the instrument is used properly, no special maintenance is required in normal operation. In some applications, product buildup on the diaphragm can influence the measuring result. Depending on the sensor and application, take precautions to ensure that heavy buildup, and especially a hardening thereof, is avoided. Cleaning If necessary, clean the diaphragm.
8 Maintenance and fault rectification Error code Cause Rectification 4 … 20 mA signal not stable Level fluctuations –– Set the integration time via the display and adjustment module or PACTware No atmospheric pres- –– Check the pressure compensation in the housing and clean the filter sure compensation element, if necessary 4 … 20 mA signal missing Connection to voltage supply wrong –– Check connection according to chapter "Connection steps" and if necessary, correct according to chapter "Wiring pl
8 Maintenance and fault rectification Fperf = √((FT)2 + (FKl)2 (+ Fa)2) With: • • • • • • Example Ftotal: Total deviation Fperf: Basic accuracy Fstab: Long-term drift FT: Temperature coefficient (influence of medium or ambient temperature) FKl: Deviation Fa: Error current output Pressure measurement in the pipeline 8 bar (800 KPa) Product temperature 50 °C, reference temperature 20 °C Calculation ΔT: ΔT = 50 °C - 20 °C = 30 K VEGABAR 54, with measuring range 25 bar, version 0.
8 Maintenance and fault rectification Total deviation analogue output signal absolute: Ftotal = 0.70 % x 8 bar/100 % = 56 mbar 8.4 Exchanging the electronics module In case of a defect, the electronics module can be exchanged by the user against an identical type. If no electronics module is available on side, the module can be ordered from the agency serving you. Ordering and exchange are possible with or without sensor serial number.
8 Maintenance and fault rectification data and displays the actual hardware and software version of the sensor. This takes approximately 60 s. Push the button "Update software" and select the previously extracted hex file. Then the software update can be started. The additional files are installed automatically. Depending on the sensor, this procedure lasts up to 1 h. Then the message appears ""Software update successfully executed". 8.
9 Dismounting 9 9.1 Dismounting Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 9.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
10 Supplement 10 Supplement 10.1 Technical data General data Parameter, pressure Gauge pressure, absolute pressure, vacuum Communication interface None Measuring principle Ceramic-capacitive, dry measuring cell Materials, weights, torque Material 316L corresponds to 1.4404 or 1.4435 Materials, wetted parts ƲƲ Process fitting ƲƲ Diaphragm 316L sapphire ceramic® (99.
10 Supplement ƲƲ Transport cover PFA Weight approx. 0.8 … 8 kg (1.764 … 17.64 lbs), depending on process fitting Max. torques ƲƲ Screws for PMC 1", PMC 1¼" 2 Nm ƲƲ G½ A, G¾ A 30 Nm ƲƲ Screws for PMC 1½" 5 Nm ƲƲ G1 A, M30 x 1.5 50 Nm ƲƲ G1" for PASVE 100 Nm Output variable Output signal 4 … 20 mA Signal resolution 1.6 µA Failure signal output current mA value unchanged 20.5 mA, 22 mA, < 3.6 mA (adjustable) Max.
10 Supplement ƲƲ Percentage value ƲƲ Pressure value -10 … 110 % -20 … 120 % Adjustment range of the zero/span adjustment relating to the nominal measuring range: ƲƲ zero -20 … +95 % ƲƲ Difference between zero and span max. 120 % of the nominal range ƲƲ Span Recommended max. turn down -120 … +120 % 12) 10 : 1 (no limitation) Nominal measuring ranges and overload capability in bar/kPa The specifications are only an overview and refer to the measuring cell.
10 Supplement Nominal range Overload capacity, max. pressure Overload capacity, min. pressure 0 … 60 bar/0 … 6000 kPa 200 bar/20 MPa 0 bar abs. Nominal measuring ranges and overload capability in psig The specifications are only an overview and refer to the measuring cell. Limitations due to the material and version of the process fitting are possible. The specifications on the nameplate apply. Nominal range Overload capacity, max. pressure Overload capacity, min. pressure 0 … +1.
10 Supplement ƲƲ Air pressure Determination of characteristics 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig) Limit point adjustment according to IEC 61298-2 Characterstic curve Linear Influence of the installation position < 0.2 mbar/20 Pa (0.
10 Supplement 0,15 % -40°C -20°C 20°C 40°C 60°C 80°C -0,15 % Fig. 20: Thermal change, current output Long-term stability (according to DIN 16086 and IEC 60770-1) Applies to digital HART interface as well as to analogue current output 4 … 20 mA under reference conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal measuring range/set span. Long-term drift of the zero signal: ƲƲ For one year < 0.05 % x TD ƲƲ For ten years < 0.2 % x TD ƲƲ For five years < 0.
10 Supplement Measuring cell seal Product temperature FFKM (Perlast G75B) -15 … +120 °C (-4 … +248 °F) Vibration resistance Shock resistance mechanical vibrations with 4 g and 5 … 100 Hz tested according to the regulations of German Lloyd, GL characteristics 2. Acceleration 100 g/6 ms15) Electromechanical data - version IP 66/IP 67 Cable entry/plug16) ƲƲ Single chamber housing –– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x blind plug M20 x 1.
10 Supplement ƲƲ Configuration four wires, one suspension wire, one breather capillary, screen braiding, metal foil, mantle ƲƲ Wire cross-section 0.5 mm² (AWG 20) ƲƲ Standard length 5 m (16.40 ft) ƲƲ Wire resistance ƲƲ Max. length ƲƲ Min. bending radius at 25 °C/77 °F ƲƲ Diameter approx. ƲƲ Colour Cable entry/plug17) ƲƲ External housing Spring-loaded terminals for wire crosssection up to < 0.036 Ω/m (0.011 Ω/ft) 180 m (590.5 ft) 25 mm (0.985 in) 8 mm (0.315 in) Blue –– 1 x cable gland M20 x 1.
10 Supplement 1000 750 2 500 1 250 3 12 14 16 18 20 22 24 26 28 30 32 34 36 V Fig.
10 Supplement Plastic housing ~ 69 mm (2.72") ø 79 mm (3.03") ~ 84 mm (3.31") M20x1,5/ ½ NPT M16x1,5 M20x1,5/ ½ NPT 1 112 mm (4.41") 112 mm (4.41") ø 79 mm (3.31") 2 Fig. 22: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Aluminium housing ~ 87 mm (3.43") ~ 116 mm (4.57") ø 86 mm (3.39") M16x1,5 120 mm (4.72") 116 mm (4.57") ø 86 mm (3.
10 Supplement Aluminium housing in protection rating IP 66/IP 68 (1 bar) ~ 105 mm (4.13") ~ 150 mm (5.91") ø 84 mm (3.31") ø 84 mm (3.31") M20x1,5 120 mm (4.72") 116 mm (4.57") M16x1,5 M20x1,5 M20x1,5/ ½ NPT 1 2 Fig. 24: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.35␣in higher 1 2 Single chamber version Double chamber version Stainless steel housing ~ 69 mm (2.72") ø 79 mm (3.11") ~ 59 mm (2.32") ø 80 mm (3.
10 Supplement Stainless steel housing in protection rating IP 66/IP 68 (1 bar) ~ 93 mm (3.66") ~ 105 mm (4.13") ø 80 mm (3.15") ~ 103 mm (4.06") ø 84 mm (3.31") ø 77 mm (3.03") 112 mm (4.41") 120 mm (4.72") 117 mm (4.61") M16x1,5 M20x1,5 M20x1,5/ ½ NPT M20x1,5/ ½ NPT 1 3 2 Fig. 26: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.
10 Supplement 68 mm (2.68") 68 mm (2.68") 42mm (1.65") 92 mm (3.62") 51 mm (2.01") 1 110 mm x 90 mm (4.33" x 3.54") 40mm (1.57") 3 2 Fig. 28: External housing - Stainless steel version 1 2 3 Lateral cable outlet Axial cable outlet Seal 2 mm (0.
10 Supplement VEGABAR 54, threaded fitting 22 mm (0.87") ø 24 mm (0.95") 21 mm (0.83") G¾A SW 27 mm (1.06") G½A 45 mm (1.77") 44 mm (1.73") SW 36 mm (1.42") GA GB GC G1A 40 mm (1.58") 21 mm (0.83") ø 30 mm (1.18") ø 40 mm (1.58") SW 32 mm (1.26") 59 mm (2.32") 44 mm (1.73") SW 41 mm (1.61") G1A ø 18 mm (0.71") ø 27 mm (1.06") ø 40 mm (1.58") GD 36725-EN-130321 Fig. 29: VEGABAR 54, threaded connection: GA = G¾ A acc.
10 Supplement 45 mm (1.77") 44,5 mm (1.75") VEGABAR 54, hygienic fitting 1 ø 36 mm (1.42") ø 50,5 mm (1.99") CI CD 67,5 mm (2.66") 44,5 mm (1.75") SW41 G1 A ø 52 mm (2.05") GK RI Fig.
10 Supplement 45 mm (1.77") VEGABAR 54, hygienic fitting 2 ø 62,5 mm (2.46") RE 36725-EN-130321 Fig. 31: VEGABAR 54, hygienic fitting: RE = bolting DN 25 PN 40 acc.
10 Supplement b d2 f 50 mm (1.97") VEGABAR 54, flange connection 1 d4 k D DB, FZ, DA, F1 1 mm DN PN D b k d2 d4 f DB 15 40 95 16 65 4 x ø14 45 2 FZ 20 40 105 18 75 4 x ø14 58 2 DA 25 40 115 18 85 4 x ø14 68 2 108 14,2 D b 79,2 4 x ø15,7 50,8 2 F1 1“ 150 lbs mm DN PN 1 DB 15 40 3.74" 0.63" 2.56" 4 x ø0,55" 1.77" 0.08" 4.13" 0.71" 2.95" 4 x ø0,55" 2.28" 4.53" 0.71" 3.35" 4 x ø0,55" 2.68" 0.08" 4.25" 0.56" 3.12" 4 x ø0,62" 0.
10 Supplement VEGABAR 54, flange connection 2 f b 47,8 mm (1 7/8") d2 d4 k D FA mm DN PN D b k d2 d4 f FA 25 40 115 18 85 4xø14 68 2 inch DN PN D b k d2 d4 FA 25 40 4 17/64" 45/64" 3 11/32" 4xø 35/64" 2 43/64" f 5/64" 36725-EN-130321 Fig.
10 Supplement VEGABAR 54, connection for paper industry 1 BF/BS M8 43,7 mm (1 23/32") 43,7 mm (1 23/32") BA/BB ø 26,3 mm (1 1/32") PC M8 ø 40,4 mm (1 19/32") ø 34,6 mm (1 23/64") PD ø 25 mm (63/64") ø 27,4 mm (1 5/64") ø 36 mm (1 27/64") M8 45,5 mm (1 51/64") M30x1,5 29 mm (1 9/64") 21 mm (53/64") M44x1,25 SW32 48 mm (1 57/64") 52 mm (2 3/64") 46 PE Fig. 34: VEGABAR 54, fitting for the paper industry: BA/BB = M44 x 1.25; BF = M30 x 1.5; BS = M30 x 1.
10 Supplement VEGABAR 54, connection for paper industry 2 SW36 G1" 30 mm (1 3/16") 104,8 mm (4 1/8") 30 mm (1 3/16") 104,8 mm (4 1/8") SW36 G1" ø 24,7 mm (31/32") VP ø 29,3 mm (1 5/32") ø 24,7 mm (31/32") VR ø 29,3 mm (1 5/32") 36725-EN-130321 Fig.
10 Supplement 10.3 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. Only in U.S.A.: Further information see patent label at the sensor housing. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.
INDEX INDEX A Accessories ––Display and adjustment module 10 ––Flanges 10 ––Measuring instrument holder 10 ––Protective cover 10 Application area 8 C Cable screening and grounding 16 Check output signal 48 Connection cable 15 Copy sensor data 30, 43 D Disposal 53 E Electronics and connection compartment 18 Error messages 49 F Fault rectification 48 H Hotline 48 I ––Electronics module 10 Reset 31, 44 Return form 52 S Safety data sheet 52 Seal concept 9 Service hotline 48 Span adjustment 30,
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2013 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.